View-angle tilting and slice-encoding metal artifact correction for artifact reduction in MRI: Experimental sequence optimization for orthopaedic tumor endoprostheses and clinical application

  • Pia M. Jungmann
  • , Carl Ganter
  • , Christoph J. Schaeffeler
  • , Jan S. Bauer
  • , Thomas Baum
  • , Reinhard Meier
  • , Mathias Nittka
  • , Florian Pohlig
  • , Hans Rechl
  • , Ruediger Von Eisenhart-Rothe
  • , Ernst J. Rummeny
  • , Klaus Woertler
  • Technical University of Munich
  • Cantonal Hospital Grisons
  • Siemens AG

Research output: Contribution to journalArticlepeer-review

34 Scopus citations

Abstract

Background: MRI plays a major role in follow-up of patients with malignant bone tumors. However, after limb salvage surgery, orthopaedic tumor endoprostheses might cause significant metal-induced susceptibility artifacts. Purposes: To evaluate the benefit of view-angle tilting (VAT) and slice-encoding metal artifact correction (SEMAC) for MRI of large-sized orthopaedic tumor endoprostheses in an experimental model and to demonstrate clinical benefits for assessment of periprosthetic soft tissue abnormalities. Methods: In an experimental setting, tumor endoprostheses (n=4) were scanned at 1.5T with three versions of optimized high-bandwidth turbo-spin-echo pulse sequences: (i) standard, (ii) VAT and (iii) combined VAT and SEMAC (VAT&SEMAC). Pulse sequences included coronal short-tau-inversion-recovery (STIR), coronal T1-weighted (w), transverse T1-w and T2-w TSE sequences. For clinical evaluation, VAT&SEMAC was compared to conventional metal artifact-reducing MR sequences (conventional MR) in n=25 patients with metal implants and clinical suspicion of tumor recurrence or infection. Diameters of artifacts were measured quantitatively. Qualitative parameters were assessed on a five-point scale (1=best, 5=worst): "image distortion", "artificial signal changes at the edges" and "diagnostic confidence". Imaging findings were correlated with pathology. T-tests and Wilcoxon-signed rank tests were used for statistical analyses. Results: The true size of the prostheses was overestimated on MRI (P<0.05). A significant reduction of artifacts was achieved by VAT (P<0.001) and VAT&SEMAC (P=0.003) compared to the standard group. Quantitative scores improved in the VAT and VAT&SEMAC group (P<0.05). On clinical MR images, artifact diameters were significantly reduced in the VAT&-SEMAC-group as compared with the conventional-group (P<0.001). Distortion and artificial signal changes were reduced and diagnostic confidence improved (P<0.05). In two cases, tumor-recurrence, in ten cases infection and in thirteen cases other pathologies were diagnosed. Conclusions: Significant reduction of metallic artifacts was achieved by VAT and SEMAC. Clinical results suggest, that these new techniques will be beneficial for detecting periprosthetic pathologies during postoperative follow-up.

Original languageEnglish
Article numbere0124922
JournalPLoS ONE
Volume10
Issue number4
DOIs
StatePublished - 24 Apr 2015

Fingerprint

Dive into the research topics of 'View-angle tilting and slice-encoding metal artifact correction for artifact reduction in MRI: Experimental sequence optimization for orthopaedic tumor endoprostheses and clinical application'. Together they form a unique fingerprint.

Cite this