Abstract
On the RT 305 apparatus for therapeutic hard Xrays the curves of the depth doses, dose profiles and factors of back scatter were measured for 8 different cones under a tube voltage of 300 kV. The measurements were made with an ionization chamber in a water phantom.
| Translated title of the contribution | Distributions of depth doses, dose profiles and factors of back scatter in 300 kV X rays |
|---|---|
| Original language | German |
| Pages (from-to) | 65-67 |
| Number of pages | 3 |
| Journal | Strahlentherapie |
| Volume | 146 |
| Issue number | 1 |
| State | Published - 1973 |