Suitability of screw plasticization for autosterile injection molding

Markus Schönberger, Franziska Hierl, Miriam Haerst, Erich Wintermantel

Research output: Contribution to journalArticlepeer-review

Abstract

Single use medical products are subject to subsequent sterilization treatment following manufacturing processes, for example, injection molding, to obtain sterility. A new approach is developed to use the injection-molding process to produce sterilized products without an additional sterilization procedure. To achieve the aim of an autosterile injection molding, it is necessary to evaluate the sterilization impact of screw plasticization regarding sterile polymer melts before entering the injection mold cavity. Within this study, a screw plasticization unit is fed with defined contaminated polymer granules. Geobacillus stearothermophilus spores are used as challenge organisms to determine the sterility grade of screw plasticization. The investigation includes method verification and the experimental testing of polyoxymethylene, polyethylene, and cyclo-olefin copolymer melts at different melt temperatures of 200°C and 280°C.

Original languageEnglish
Article number21403
JournalAdvances in Polymer Technology
Volume33
Issue number2
DOIs
StatePublished - 2014

Keywords

  • Autosterile
  • Injection molding
  • Medical
  • Polymer melt

Fingerprint

Dive into the research topics of 'Suitability of screw plasticization for autosterile injection molding'. Together they form a unique fingerprint.

Cite this