Spatial resolution characterization of a X-ray microCT system

J. Rueckel, M. Stockmar, F. Pfeiffer, J. Herzen

Research output: Contribution to journalArticlepeer-review

76 Scopus citations

Abstract

We report on an experimental characterization of the spatial resolution of a commercial X-ray micro-computed tomography scanner. We have measured the full modulation transfer function (MTF) to assess the spatial resolution. The MTF and those spatial frequencies corresponding to a contrast loss of 50% were determined as a function of different applied X-ray tube parameters and magnification-dependent pixel sizes. A significant influence of the focal spot enlargement on the achievable spatial resolution could be shown. Our results allow for the designation of optimal X-ray tube parameters for a specific application requirement.

Original languageEnglish
Pages (from-to)230-234
Number of pages5
JournalApplied Radiation and Isotopes
Volume94
DOIs
StatePublished - 1 Sep 2014

Keywords

  • General electric
  • MTF
  • Micro-CT
  • Spatial resolution
  • V|tome|xs
  • X-ray imaging

Fingerprint

Dive into the research topics of 'Spatial resolution characterization of a X-ray microCT system'. Together they form a unique fingerprint.