Simultaneous measurement of cellular respiration and acidification with a single CMOS ISFET

  • Mirko Lehmann
  • , Werner Baumann
  • , Martin Brischwein
  • , Hans Jürgen Gahle
  • , Ingo Freund
  • , Ralf Ehret
  • , Sabine Drechsler
  • , Heiko Palzer
  • , Martin Kleintges
  • , Ullrich Sieben
  • , Bernhard Wolf

Research output: Contribution to journalArticlepeer-review

117 Scopus citations

Abstract

In vivo, the pH value and oxygen partial pressure are the most important physico-chemical parameters in the microenvironment of human tissues. In vitro, the extracellular acidification rate of cell cultures is an indicator of global cellular metabolism, while the rate of oxygen consumption is a measure of mitochondrial activity. Earlier approaches had the disadvantage that these two values had to be measured with two separate sensors at different loci within the tissue or cell culture. Furthermore, conventional Clark-type oxygen sensors are not very compatible for miniaturisation, making it impossible to measure at small cell volumes or even at the single cell level. We have, therefore, developed an ISFET based sensor structure which is able to measure both pH and oxygen partial pressure. This sensor structure was tested in vitro for simultaneous records of cellular acidification and respiration rates at the same site within the cell culture. This sensor is manufactured by a CMOS-process.

Original languageEnglish
Pages (from-to)195-203
Number of pages9
JournalBiosensors and Bioelectronics
Volume16
Issue number3
DOIs
StatePublished - 1 May 2001

Keywords

  • Acidification
  • Cellular sensor
  • Heterogeneity
  • ISFET
  • Respiration

Fingerprint

Dive into the research topics of 'Simultaneous measurement of cellular respiration and acidification with a single CMOS ISFET'. Together they form a unique fingerprint.

Cite this