Abstract
Purpose of this study was to evaluate the reliability and validity of 4D rasterstereography under dynamic conditions. Therefore simulated anatomical fixed points on a wooden plate were measured during different movements. Seven different motion patterns in all three angles of space were evaluated. The simulated parameter trunk length was measured with an accuracy of 3.58. mm (SD±3.29. mm) and the dimple distance was detected with an accuracy of 0.88. mm (SD±1.04. mm). With this rasterstereographic prototype it is possible to examine dynamically the spinal posture with adequate accuracy.
Original language | English |
---|---|
Pages (from-to) | 308-312 |
Number of pages | 5 |
Journal | Computers in Biology and Medicine |
Volume | 41 |
Issue number | 6 |
DOIs | |
State | Published - Jun 2011 |
Externally published | Yes |
Keywords
- Motion study
- Rasterstereography
- Reliability
- Spine
- Validity