Region of interest focused MRI to synthetic CT translation using regression and segmentation multi-task network

Sandeep S. Kaushik, Mikael Bylund, Cristina Cozzini, Dattesh Shanbhag, Steven F. Petit, Jonathan J. Wyatt, Marion I. Menzel, Carolin Pirkl, Bhairav Mehta, Vikas Chauhan, Kesavadas Chandrasekharan, Joakim Jonsson, Tufve Nyholm, Florian Wiesinger, Bjoern Menze

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Objective. In MR-only clinical workflow, replacing CT with MR image is of advantage for workflow efficiency and reduces radiation to the patient. An important step required to eliminate CT scan from the workflow is to generate the information provided by CT via an MR image. In this work, we aim to demonstrate a method to generate accurate synthetic CT (sCT) from an MR image to suit the radiation therapy (RT) treatment planning workflow. We show the feasibility of the method and make way for a broader clinical evaluation. Approach. We present a machine learning method for sCT generation from zero-echo-time (ZTE) MRI aimed at structural and quantitative accuracies of the image, with a particular focus on the accurate bone density value prediction. The misestimation of bone density in the radiation path could lead to unintended dose delivery to the target volume and results in suboptimal treatment outcome. We propose a loss function that favors a spatially sparse bone region in the image. We harness the ability of the multi-task network to produce correlated outputs as a framework to enable localization of region of interest (RoI) via segmentation, emphasize regression of values within RoI and still retain the overall accuracy via global regression. The network is optimized by a composite loss function that combines a dedicated loss from each task. Main results. We have included 54 brain patient images in this study and tested the sCT images against reference CT on a subset of 20 cases. A pilot dose evaluation was performed on 9 of the 20 test cases to demonstrate the viability of the generated sCT in RT planning. The average quantitative metrics produced by the proposed method over the test set were—(a) mean absolute error (MAE) of 70 ± 8.6 HU; (b) peak signal-to-noise ratio (PSNR) of 29.4 ± 2.8 dB; structural similarity metric (SSIM) of 0.95 ± 0.02; and (d) Dice coefficient of the body region of 0.984 ± 0. Significance. We demonstrate that the proposed method generates sCT images that resemble visual characteristics of a real CT image and has a quantitative accuracy that suits RT dose planning application. We compare the dose calculation from the proposed sCT and the real CT in a radiation therapy treatment planning setup and show that sCT based planning falls within 0.5% target dose error. The method presented here with an initial dose evaluation makes an encouraging precursor to a broader clinical evaluation of sCT based RT planning on different anatomical regions.

Original languageEnglish
Article number195003
JournalPhysics in Medicine and Biology
Volume68
Issue number19
DOIs
StatePublished - 7 Oct 2023
Externally publishedYes

Keywords

  • MRI radiation therapy
  • focused loss