Rapid single-scan T2*-mapping using exponential excitation pulses and image-based correction for linear background gradients

  • Simon Baudrexel
  • , Steffen Volz
  • , Christine Preibisch
  • , Johannes C. Klein
  • , Helmuth Steinmetz
  • , Rüdiger Hilker
  • , Ralf Deichmann

Research output: Contribution to journalArticlepeer-review

68 Scopus citations

Abstract

A method for fast quantitative T2* mapping based on multiple gradient-echo (multi-GE) imaging with correction for static magnetic field inhomogeneities is described, using an exponential excitation pulse. Field gradient maps are obtained from the phase information and modulus data are subsequently corrected, allowing for simple monoexponential T2* fitting. Echoes with long echo times suffering from major signal losses due to field inhomogeneities are excluded from the analysis. The acquisition time for a matrix size of 256 x 256, 1 mm in-plane resolution, and 2 mm slice thickness amounts to 15 s per slice. An additional correction for in-plane field gradients further improves accuracy. Phantom experiments show that the method provides accurate T2* values for field gradients up to 200 μT/m; for gradients up to 300 μT/m errors do not exceed 15%. In vivo T 2* values acquired on healthy volunteers at 3T are in excellent agreement with results from the literature.

Original languageEnglish
Pages (from-to)263-268
Number of pages6
JournalMagnetic Resonance in Medicine
Volume62
Issue number1
DOIs
StatePublished - Jul 2009
Externally publishedYes

Keywords

  • Effective transverse relaxation time
  • Field inhomogeneity
  • Gradient echo
  • Susceptibility artifact
  • T*

Fingerprint

Dive into the research topics of 'Rapid single-scan T2*-mapping using exponential excitation pulses and image-based correction for linear background gradients'. Together they form a unique fingerprint.

Cite this