Quantitative determination of phosgene doses by reflectometric badge readout

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

Commercial phosgene dosimeter badges are lacking precise and sensitive analysis when used only by visual comparison to a color reference. To meet the discussed occupational standard set to 54 ppm min, objective quantification by reflectance measurement is proposed. At 573 nm, the pink dye ("Koeniǵs salt") formed at the membrane surface by reaction of phosgene, and aromatic amine shows a strict linear relationship in reflectance between doses of 10 and 300 ppm min. The detection limit is calculated to 29 ppm min.

Original languageEnglish
Pages (from-to)2285-2288
Number of pages4
JournalAnalytical and Bioanalytical Chemistry
Volume397
Issue number6
DOIs
StatePublished - Jul 2010

Keywords

  • Dosimetry
  • Phosgene
  • Reflectance

Fingerprint

Dive into the research topics of 'Quantitative determination of phosgene doses by reflectometric badge readout'. Together they form a unique fingerprint.