Skip to main navigation Skip to search Skip to main content

Preclinical evaluation of a novel polyphosphazene surface modified stent. Koppara Preclinical testing of a PzF-coated stent

  • Tobias Koppara
  • , Kenichi Sakakura
  • , Erica Pacheco
  • , Qi Cheng
  • , Xiao Qing Zhao
  • , Eduardo Acampado
  • , Aloke V. Finn
  • , Mark Barakat
  • , Luc Maillard
  • , Jane Ren
  • , Mahesh Deshpande
  • , Frank D. Kolodgie
  • , Michael Joner
  • , Renu Virmani
  • CVPath Institute, Inc.
  • CeloNova BioSciences, Inc.
  • Service de Cardiologie

Research output: Contribution to journalArticlepeer-review

28 Scopus citations

Abstract

Background Treatment options for patients with coronary artery disease at high risk for bleeding complications are limited. The aim of the current preclinical study was to evaluate neointimal coverage, endothelial recovery, inflammation and thrombogenicity in a novel thin-strut (71 μm thickness) Cobalt Chromium (CoCr) stent modified with a nano-thin Polyzene®-F (PzF) surface coating. Methods and results Twenty-eight single PzF nano-coated stents and 20 bare metal control stents (BMS) were implanted in the coronary arteries of 24 pigs, with scheduled 5- (n = 5), 28- (n = 13), and 90-day (n = 6) follow-up in addition to overlapping configuration (n = 6 each), examined at 28-days. Histomorphometric analysis showed significantly lower neointimal thickness in PzF nano-coated stents than BMS controls at both 28- and 90-days (p = 0.023 and 0.005) and reduced inflammation (p = 0.06 and 0.13). Endothelial coverage over luminal surfaces at all time points was similar between nano-coated stents and BMS controls. We conducted supplementary in-vitro experiments using human monocytes and an ex-vivo swine carotid-jugular arterio-venous shunt model to better understand the healing properties afforded by the PzF nano-coating. Overall, the PzF-nano-coating showed reduced monocyte adhesion and thrombus formation compared to the un-coated controls. Conclusions Stents modified with a nano-thin PzF-coating implanted in healthy swine indicate favorable vascular healing properties shown by reduced neointimal hyperplasia and inflammation, along with resistance to thrombus formation in an ex-vivo shunt model over unmodified stents.

Original languageEnglish
Pages (from-to)217-225
Number of pages9
JournalInternational Journal of Cardiology
Volume222
DOIs
StatePublished - 1 Nov 2016
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Bare metal stent
  • Coronary artery disease
  • Percutaneous coronary intervention

Fingerprint

Dive into the research topics of 'Preclinical evaluation of a novel polyphosphazene surface modified stent. Koppara Preclinical testing of a PzF-coated stent'. Together they form a unique fingerprint.

Cite this