Skip to main navigation Skip to search Skip to main content

Plate fit of a novel clavicula plate system (VariAx Clavicle) developed using a CT-based anatomical bone database compared to reconstruction plates individually formed by surgeons

Translated title of the contribution: Plate fit of a novel clavicula plate system (VariAx Clavicle) developed using a CT-based anatomical bone database compared to reconstruction plates individually formed by surgeons
  • Benjamin Kienast
  • , Rainer H. Burgkart
  • , Andreas Petersik
  • , Nils Reimers
  • , Hanjo Neumann
  • , Claudia Beimel
  • , Arndt P. Schulz
  • Berufsgenossenschaftliches Unfallkrankenhaus Hamburg
  • Universitätsklinikum Schleswig-Holstein Campus Lübeck
  • Stryker Trauma GmbH

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Background: The aim of this study was the comparison of the plate to bone fit of manually bent reconstruction plates with two industrial prebent clavicle midshaft plates (VariAx Clavicle, Stryker, Duisburg, Germany). The precontoured plates were developed using an anatomical 3D computed tomography (CT)-bone database. Methods: From an anatomical database of 334 clavicles 8 were randomly chosen and manufactured from artificial bone as physical 3D models using a rapid prototyping system and three surgeons of various qualification levels performed manual bending. The plate to bone fit of the manually bent plates was then compared to the fit of the VariAx midshaft plates. Results: The industrially manufactured precontoured plates showed a high rate of plate fit with a fit error of less than 1.5 mm2 and with a significant difference in comparison to the manually bent plates. Conclusions: The development of osteosynthesis plates using anatomic 3D databases led to a higher degree of plate fit when compared to manually bent plates even for bones with a variable structure, such as the clavicle.

Translated title of the contributionPlate fit of a novel clavicula plate system (VariAx Clavicle) developed using a CT-based anatomical bone database compared to reconstruction plates individually formed by surgeons
Original languageEnglish
Pages (from-to)226-230
Number of pages5
JournalObere Extremitat
Volume8
Issue number4
DOIs
StatePublished - Dec 2013

Keywords

  • Automated computer based plate fit
  • Bending
  • Bone database
  • Clavicle
  • Osteosynthesis plate

Fingerprint

Dive into the research topics of 'Plate fit of a novel clavicula plate system (VariAx Clavicle) developed using a CT-based anatomical bone database compared to reconstruction plates individually formed by surgeons'. Together they form a unique fingerprint.

Cite this