Patient eligibility for application of a two-filter cerebral embolic protection device during transcatheter aortic valve implantation: Does one size fit all?

Stephanie Voss, Johanna Schechtl, Christian Nöbauer, Sabine Bleiziffer, Rüdiger Lange

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

OBJECTIVES: This study sought to determine the percentage of patients potentially eligible for implantation of the SentinelTM Cerebral Protection System (Sentinel-CPS) during transcatheter aortic valve implantation (TAVI) and to identify the reasons for treatment exclusion. METHODS: We retrospectively performed an analysis of pre-TAVI multislice computed tomography (MSCT) aortograms and data review of all patients undergoing a TAVI procedure in 2017 (n = 317). MSCT evaluation included the assessment of aortic arch anatomy and the vascular dimensions of the brachiocephalic and left common carotid artery. Data analysis focused on comorbid conditions, precluding 6-Fr sheath radial access and filter deployment due to history of previous artery interventions. RESULTS: MSCT and data analysis showed Sentinel-CPS compatibility in 61.5% of patients (n = 195). Sentinel-CPS would have been contraindicated in 38.5% (n = 122) due to one or more of the following: (i) measured diameters of the filter-landing zones <9 or >15 mm in the brachiocephalic artery and <6.5 or >10 mm in the left common carotid artery (n = 116; 88 with carotid dimensions too small); (ii) significant subclavian artery stenosis (n = 4) or an aberrant subclavian artery (n = 3) precluding Sentinel-CPS implantation and (iii) clinical characteristics including hypersensitivity to nickel-titanium (n = 1), radial artery occlusion (n = 1) or previous left common carotid artery interventions (n = 5). CONCLUSIONS: MSCT and clinical data supported Sentinel-CPS compatibility in 61.5% of patients. The most common reason for treatment exclusion was inappropriate diameter within the target landing zone of the left carotid artery. Future device development should address this limitation.

Original languageEnglish
Pages (from-to)605-612
Number of pages8
JournalInteractive Cardiovascular and Thoracic Surgery
Volume30
Issue number4
DOIs
StatePublished - 2021

Keywords

  • Cerebral embolic protection
  • Dual-filter system
  • Transcatheter aortic valve implantation

Fingerprint

Dive into the research topics of 'Patient eligibility for application of a two-filter cerebral embolic protection device during transcatheter aortic valve implantation: Does one size fit all?'. Together they form a unique fingerprint.

Cite this