PALM — Precision Attenuation Length Measurement in liquid scintillators

Vincent Rompel, Sabrina Franke, Florian Kübelbäck, Lothar Oberauer, Luca Schweizer, Korbinian Stangler, Hans Th J. Steiger, Matthias Raphael Stock, Andreas Ulrich

Research output: Contribution to journalArticlepeer-review

Abstract

Future neutrino experiments at low energies such as JUNO or Theia will use large volume homogeneous liquid scintillator detectors. The optical attenuation length of the liquid is of uttermost importance for the successful realization of these experiments. At TU Munich a new optical spectrometer (Precision Attenuation Length Measurement (PALM)) has been set up in order to measure the light attenuation of liquids which can be used for these types of experiments up to around 100 m in the wavelength region between 400 nm and 1000 nm. The setup features an optical imaging system with a long focal length, which allows for in-situ monitoring of the beam stability during the measurement. The setup capability as well as the reproducibility of its results has been demonstrated at two different wavelengths of 430 nm and 500 nm.

Original languageEnglish
Article numberP04019
JournalJournal of Instrumentation
Volume20
Issue number4
DOIs
StatePublished - 1 Apr 2025

Keywords

  • Neutrino detectors
  • Scintillators, scintillation and light emission processes (solid, gas and liquid scintillators)

Fingerprint

Dive into the research topics of 'PALM — Precision Attenuation Length Measurement in liquid scintillators'. Together they form a unique fingerprint.

Cite this