Abstract
Purpose: The MR image quality after intracranial aneurysm clipping is often impaired because of artifacts induced by metal implants. The purpose of the present study was to evaluate the benefit of a new WARP sequence with slice-encoding metal artifact correction (SEMAC) and view-angle tilting (VAT) MR imaging as novel artifact reduction techniques. Materials and Methods: A new WARP TSE (a work-in-progress software package provided by Siemens Healthcare) sequence was implemented for cranial applications based on a turbo spin echo (TSE) sequence. T1- and T2-weighted images with standard and WARP TSE sequences were acquired from 6 patients with 11 clipping sites, and the images were compared based on artifact size and general image quality. Results: T2- and T1-weighted WARP TSE sequences resulted in a highly significant reduction of metal artifacts compared with standard sequences (T2w- WARP TSE: 89.8 ± 1.4 %; T1w- WARP TSE: 84.9 ± 2.9 %; p < 0.001) without a substantial loss of image quality. Conclusion: The use of a new WARP TSE sequence after aneurysm clipping is highly beneficial for increasing the diagnostic MR image quality due to a striking reduction of metal artifacts.
Original language | English |
---|---|
Pages (from-to) | 31-37 |
Number of pages | 7 |
Journal | Clinical Neuroradiology |
Volume | 26 |
Issue number | 1 |
DOIs | |
State | Published - 1 Mar 2016 |
Keywords
- Artifact
- Clip
- MRI
- SEMAC
- WARP