Abstract
Background: The purpose of this study was to examine biomechanical properties and the degree of radiolucency of two cemented basic glenoid designs for total shoulder arthroplasty. Our hypothesis was that a component with increased micro-motion in the laboratory at time zero would also exhibit a greater amount of radiolucency in patients at a minimum of 2 years post total shoulder arthroplasty. Methods: Thirty cadaveric shoulders were divided into 2 groups (keel vs. peg). The glenoid components were first loaded with a single axial eccentric force of 196 N in all orientations and then with a transversal load of 49 N to simulate in vivo loads with abduction. Displacement of the glenoid component was determined with four different linear variable-differential transducers. In the second phase, 56 antero-posterior x-rays of 52 patients with either the same keeled (n = 24) or pegged (n = 32) glenoid component with a minimum of 24 months follow-up were evaluated for radiolucency. Results: Biomechanically the pegged glenoid showed a significant increase in micro-motion during eccentric axial loading as well as during combined loading in the anterior, posterior, and inferior position as compared to the keeled glenoid (p < 0.05). In contrast all results were significant with greater radiolucency for the keeled glenoid component (p = 0.001). Conclusion: While the pegged component exhibited a greater amount of micro-motion during biomechanical testing, radiolucency was greater in patients with a keeled component. These findings provide support for both components from different perspectives and highlight the need for well-constructed studies to determine whether glenoid design has an effect on clinical outcome, because influences are multifactorial and biomechanical forces may not recreate forces seen in vivo.
| Original language | English |
|---|---|
| Article number | 213 |
| Journal | BMC Musculoskeletal Disorders |
| Volume | 18 |
| Issue number | 1 |
| DOIs | |
| State | Published - 25 May 2017 |
| Externally published | Yes |
Keywords
- Biomechanics
- Keel glenoid
- Peg glenoid
- Shoulder prosthesis
Fingerprint
Dive into the research topics of 'No correlation between radiolucency and biomechanical stability of keeled and pegged glenoid components'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver