Skip to main navigation Skip to search Skip to main content

Neutron tomography using an elliptic focusing guide

  • N. Kardjilov
  • , A. Hilger
  • , M. Dawson
  • , I. Manke
  • , J. Banhart
  • , M. Strobl
  • , P. Böni
  • University of Salford
  • School of Computing, Science and Engineering
  • Technische Universität Berlin
  • Heidelberg University
  • Hahn-Meitner-Platz 1

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

An elliptic focusing neutron guide was combined with a cold neutron imaging instrument to produce a cone-beam, whose parameters were then characterized both at the focal point and at the detector position. This point source geometry provides improved experimental conditions for tomographic applications, providing a larger beam cross-section and enabling geometric magnification. The experimental data were compared with Monte Carlo simulations and both experiments and simulations demonstrate superior spectral and geometric homogeneity of the cone-beam setup compared to the conventional pinhole geometry.

Original languageEnglish
Article number034905
JournalJournal of Applied Physics
Volume108
Issue number3
DOIs
StatePublished - 1 Aug 2010

Fingerprint

Dive into the research topics of 'Neutron tomography using an elliptic focusing guide'. Together they form a unique fingerprint.

Cite this