Magnetic resonance real-time imaging for the evaluation of left ventricular function

Eike Nagel, Uta Schneider, Simon Schalla, Tareq Ibrahim, Bernhard Schnackenburg, Axel Bornstedt, Christoph Klein, Hans B. Lehmkuhl, Eckart Fleck

Research output: Contribution to journalArticlepeer-review

56 Scopus citations

Abstract

New ultrafast gradient systems and hybrid imaging sequences make it possible to acquire a complete image in real time, without the need for breathholding or electrocardiogram (ECG) triggering. In 21 patients, left ventricular function was assessed by the use of a turbo-gradient echo technique, an echo-planar imaging (EPI) technique, and a new real-time imaging technique. End-diastolic and end-systolic volumes, left ventricular muscle mass, and ejection fraction of the ultrafast techniques were compared with the turbo-gradient echo technique. Inter- and intraobserver variability was determined for each technique. Image quality was sufficient for automated contour detection in all but two patients in whom foldover occurred in the real-time images. Results of the ultrafast imaging techniques were comparable with conventional turbo-gradient echo techniques. There was a tendency to overestimate the end-diastolic volume by 3.9 and 1.3 ml with EPI real-time imaging, the end-systolic volume by 0.9 and 5.0 ml, and the left ventricular mass by 2.6 and 23.8 g. Ejection fraction showed a tendency to be overestimated by 1.1% with EPI and underestimated by 4.5% with real-time imaging. Correlation between EPI real-time imaging and turbo-gradient echo were 0.94 and 0.95, respectively, for end-diastolic volumes, 0.98 and 0.96, respectively, for end-systolic volumes, and 0.96 and 0.89, respectively, for left ventricular mass. Inter- and intraobserver variability was low with all three techniques. Real-time imaging allows an accurate determination of left ventricular function without ECG triggering. Scan times can be reduced significantly with this new technique. Further studies will have to assess the value of real-time imaging for the detection of wall motion abnormalities and the imaging of patients with atrial fibrillation.

Original languageEnglish
Pages (from-to)7-14
Number of pages8
JournalJournal of Cardiovascular Magnetic Resonance
Volume2
Issue number1
DOIs
StatePublished - 2000
Externally publishedYes

Keywords

  • Left ventricular function
  • Magnetic resonance
  • Real-time imaging

Fingerprint

Dive into the research topics of 'Magnetic resonance real-time imaging for the evaluation of left ventricular function'. Together they form a unique fingerprint.

Cite this