Indocyanine green matching phantom for fluorescence-guided surgery imaging system characterization and performance assessment

Alberto J. Ruiz, Mindy Wu, Ethan P.M. LaRochelle, Dimitris Gorpas, Vasilis Ntziachristos, T. Joshua Pfefer, Brian W. Pogue

Research output: Contribution to journalArticlepeer-review

37 Scopus citations

Abstract

SIGNIFICANCE: Expanded use of fluorescence-guided surgery with devices approved for use with indocyanine green (ICG) has led to a range of commercial systems available. There is a compelling need to be able to independently characterize system performance and allow for cross-system comparisons. AIM: The goal of this work is to expand on previous proposed fluorescence imaging standard designs to develop a long-term stable phantom that spectrally matches ICG characteristics and utilizes 3D printing technology for incorporating tissue-equivalent materials. APPROACH: A batch of test targets was created to assess ICG concentration sensitivity in the 0.3- to 1000-nM range, tissue-equivalent depth sensitivity down to 6 mm, and spatial resolution with a USAF test chart. Comparisons were completed with a range of systems that have significantly different imaging capabilities and applications, including the Li-Cor® Odyssey, Li-Cor® Pearl, PerkinElmer® Solaris, and Stryker® Spy Elite. RESULTS: Imaging of the ICG-matching phantoms with all four commercially available systems showed the ability to benchmark system performance and allow for cross-system comparisons. The fluorescence tests were able to assess differences in the detectable concentrations of ICG with sensitivity differences >10× for preclinical and clinical systems. Furthermore, the tests successfully assessed system differences in the depth-signal decay rate, as well as resolution performance and image artifacts. The manufacturing variations, photostability, and mechanical design of the tests showed promise in providing long-term stable standards for fluorescence imaging. CONCLUSIONS: The presented ICG-matching phantom provides a major step toward standardizing performance characterization and cross-system comparisons for devices approved for use with ICG. The developed hybrid manufacturing platform can incorporate long-term stable fluorescing agents with 3D printed tissue-equivalent material. Further, long-term testing of the phantom and refinements to the manufacturing process are necessary for future implementation as a widely adopted fluorescence imaging standard.

Original languageEnglish
Pages (from-to)1-15
Number of pages15
JournalJournal of Biomedical Optics
Volume25
Issue number5
DOIs
StatePublished - 1 May 2020

Keywords

  • fluorescence-guided surgery
  • imaging standard
  • indocyanine green
  • surgery
  • tissue simulating phantoms

Fingerprint

Dive into the research topics of 'Indocyanine green matching phantom for fluorescence-guided surgery imaging system characterization and performance assessment'. Together they form a unique fingerprint.

Cite this