Increasing isoflurane concentration may cause paradoxical increases in the EEG bispectral index in surgical patients

O. Detsch, G. Schneider, E. Kochs, G. Hapfelmeier, C. Werner

Research output: Contribution to journalArticlepeer-review

76 Scopus citations

Abstract

We have studied the effects of increases in isoflurane concentration on the EEG bispectral index (BIS) in 70 patients anaesthetized with isoflurane-nitrous oxide-sufentanil for major abdominal surgery. During surgery, baseline BIS was recorded at 0.8% end-tidal isoflurane with nitrous oxide in oxygen (FI(O2) 0.35). After this, end-tidal isoflurane was increased to 1.6% for 15 min and decreased subsequently to 0.8% for 20 min to assess recovery. In 20 patients, BIS decreased from a mean value of 40 (SD 9) during baseline to 25 (10) at 1.6% isoflurane. In contrast, BIS did not change in 23 patients and increased in 27 patients from 35 (6) to 46 (8) as isoflurane was increased to 1.6%. In all patients, BIS recovered to baseline values at 0.8% isoflurane. The changes in BIS with increasing isoflurane concentration were not related to drugs or differences in physiological variables, which did not differ between groups. Patients with a decrease in BIS were significantly younger (38 (range 18-68) yr) than those with unchanged (55 (26-70) yr) or increased (60 (40-70) yr) BIS values (P < 0.001). It is possible that the paradoxical increase in BIS is related to continuous pre-burst EEG patterns consisting of high-frequency activity. This suggests that the use of BIS as a guide for isoflurane administration may be misleading in some patients undergoing surgical procedures.

Original languageEnglish
Pages (from-to)33-37
Number of pages5
JournalBritish Journal of Anaesthesia
Volume84
Issue number1
DOIs
StatePublished - Jan 2000

Keywords

  • Anaesthesia, depth
  • Anaesthetic techniques, inhalation
  • Anaesthetics volatile, isoflurane
  • Monitoring, bispectral index
  • Monitoring, electroencephalography

Fingerprint

Dive into the research topics of 'Increasing isoflurane concentration may cause paradoxical increases in the EEG bispectral index in surgical patients'. Together they form a unique fingerprint.

Cite this