Flow-injection electrochemical immunosensor for the detection of human IgG based on glucose oxidase-derivated biomimetic interface

Research output: Contribution to journalArticlepeer-review

67 Scopus citations

Abstract

A newly flow-through electrochemical immunosensor for monitoring IgG in human serum has been developed by using core-shell SiO2/Au nanocomposites and poly(amidoamine) G4 dendrimer as matrices. The ferrocenecarboaldehyde-labeled anti-IgG biomolecules were initially chemisorbed onto the nanoparticle surface, and then glucose oxidase (GOx), as a blocking reagent instead of bovine serum albumin (BSA), was backfilled onto the modified surface. The formation of the antibody-antigen complex by a simple one-step immunoreaction between the immobilized anti-IgG and IgG in sample solution introduced a barrier of direct electrical communication between the immobilized GOx and the base surface, and decreased the immobilized GOx toward the catalytic oxidation of glucose. The performance and factors influencing the performance of the immunosensor were evaluated. Under optimal conditions, the linear range of the developed immunosensor by using GOx as enhancer was from 5.0 × 10-6 to 9.6 × 10-4 mol/L with a detection limit of 8.0 × 10-7 mol/L IgG (at 3δ), while the detection limit by using BSA was 1.5 × 10-5 mol/L IgG (at 3δ) with the linear range from 3.5 × 10-5 to 1.2 × 10-3 mol/L. The selectivity, reproducibility and stability of the proposed immunosensor were acceptable. The IgG contents in 37 human serum samples obtained by the proposed method are identical with the data of clinical laboratory.

Original languageEnglish
Pages (from-to)2125-2130
Number of pages6
JournalBiosensors and Bioelectronics
Volume24
Issue number7
DOIs
StatePublished - 15 Mar 2009

Keywords

  • Biomimetic interface
  • Bovine serum albumin
  • Electrochemical immunosensor
  • Glucose oxidase
  • Human IgG

Fingerprint

Dive into the research topics of 'Flow-injection electrochemical immunosensor for the detection of human IgG based on glucose oxidase-derivated biomimetic interface'. Together they form a unique fingerprint.

Cite this