Abstract
The conventional thermoresistive sensor-based feedback constant temperature circuit have shown some performance limitations due to the input offset voltage of the amplifier. The d.c. analysis of this circuit has been presented to graphically demonstrate these limitations. Alternative feedback measurement scheme without employing the Wheatstone bridge is proposed. PI and predictive controller designs are described. Simulation results for these controllers and a practical configuration are presented.
Original language | English |
---|---|
Pages | 1062-1066 |
Number of pages | 5 |
State | Published - 1998 |
Externally published | Yes |
Event | Proceedings of the 1998 IEEE Instrumentation and Measurement Technology Conference, IMTC. Part 1 (of 2) - St.Paul, MN, USA Duration: 18 May 1998 → 21 May 1998 |
Conference
Conference | Proceedings of the 1998 IEEE Instrumentation and Measurement Technology Conference, IMTC. Part 1 (of 2) |
---|---|
City | St.Paul, MN, USA |
Period | 18/05/98 → 21/05/98 |