Abstract
Endoscopy is gradually replacing open surgery in the gastrointestinal tract. Therefore, novel medical devices and instrumentation are required, such as flexible miniaturized mechanisms for tissue joining and manipulation. In this paper, an absorbable implant for the purpose of long-term tissue fixation is presented. An experimental validation of the implant design and functionality is introduced. The implant achieves tissue penetration and provides flexible adaptation according to the thickness of two stomach walls. This mechanism is easy as it is based on pushpull principle using unidirectional forces. The shape optimization of each implant part occurs by varying design-influencing factors. The load transmission on postmortem porcine tissue was measured in the frame of the experimental setup. The feasibility of the implant was tested, and the forces needed for the intended application quantified. The implant successfully achieves tissue penetration, load transmission, adjustment, and fixation. It is a new alternative to conventional tissue-joining mechanisms.
Original language | English |
---|---|
Article number | 5604659 |
Pages (from-to) | 429-434 |
Number of pages | 6 |
Journal | IEEE Transactions on Biomedical Engineering |
Volume | 58 |
Issue number | 2 |
DOIs | |
State | Published - Feb 2011 |
Keywords
- Absorbable implant
- gastrointestinal endoscopy
- stomach
- tissue joining