Abstract
This paper presents the first experimental results on magnetic particle imaging with full 2D encoding. The encoding speed achieved was 3.88 ms for a field of view of 1 × 1 cm2. Small phantoms composed of several dots each filled with 200 nl undiluted Resovist (500 mmol(Fe) l-1) were scanned. A resolution of better than 1 mm was achieved for a frame rate of 25 frames s-1.
Original language | English |
---|---|
Pages (from-to) | N81-N84 |
Journal | Physics in Medicine and Biology |
Volume | 53 |
Issue number | 6 |
DOIs | |
State | Published - 21 Mar 2008 |