T1 - Evaluation of feasibility and image quality of 68Ga-DOTATOC positron emission tomography/magnetic resonance in comparison with positron emission tomography/computed tomography in patients with neuroendocrine tumors
AU - Gaertner, Florian C.
AU - Beer, Ambros J.
AU - Souvatzoglou, Michael
AU - Eiber, Matthias
AU - Fürst, Sebastian
AU - Ziegler, Sybille I.
AU - Brohl, Florian
AU - Schwaiger, Markus
AU - Scheidhauer, Klemens
PY - 2013/5
Y1 - 2013/5
N2 - OBJECTIVES: The primary aims of this study were to evaluate the feasibility of simultaneous (DOTA-Phe-Tyr)octreotide positron emission tomography (PET)/magnetic resonance (MR) acquisition on a fully integrated PET/MR scanner in patients and to compare the quality of PET images acquired with a PET/MR device with those acquired with a PET/computed tomography (CT) scanner. PATIENTS AND METHODS: Sequential PET/CT and PET/MR imaging was performed in 24 patients with neuroendocrine tumors using a single-injection/dual-imaging protocol. After intravenous injection of Ga-DOTATOC (mean, 120 MBq), PET/CT imaging including low-dose CT was performed at a mean time of 17 minutes post injection, and subsequently, PET/MR imaging including a Dixon sequence for attenuation correction was started at a mean time of 82 minutes post injection. The PET/CT and PET/MR images were analyzed visually using a 4-point scale for quality, coregistration, anatomical correlation, and lesion conspicuity. The standardized uptake value of background organs and focal lesions was measured and compared between the PET/CT and PET/MR acquisitions. RESULTS: Ga-DOTATOC PET acquired on the PET/MR delivered images with a good diagnostic quality (average visual rating PET/CT, 2.83; PET/MR, 2.08; P <; 0.01). The standardized uptake value of focal lesions did not differ between the PET/CT and PET/MR acquisitions (P >; 0.3) and correlated in a linear fashion (correlation coefficient ρ = 0.90). Lesion conspicuity was slightly, but significantly, higher on the PET/CT acquisitions (PET/CT, 2.71; PET/MR, 2.62; P = 0.01). Positron emission tomography/MR detected 153 of 157 lesions identified by PET/CT; however, there was no difference in sensitivity on a patient basis or organ system basis. Anatomical correlates for focal PET lesions could significantly more often be delineated using MR Dixon images compared with low-dose CT (average visual rating PET/CT, 1.78; PET/MR, 2.30; P <; 0.01). Coregistration of functional and morphological data was better on PET/MR compared with PET/CT, which, however, did not reach significance (average visual rating PET/CT, 2.17; PET/MR, 2.46; P = 0.10). CONCLUSIONS: Ga-DOTATOC PET/MR imaging is feasible in patients, with good image quality, and detectability of focal PET lesions was equivalent to PET/CT on a patient basis and organ system basis. Now, the clinical value of Ga-DOTATOC PET/MR with additional diagnostic MR protocols has to be evaluated against PET/CT with multiphase contrast-enhanced CT protocols in future studies.
AB - OBJECTIVES: The primary aims of this study were to evaluate the feasibility of simultaneous (DOTA-Phe-Tyr)octreotide positron emission tomography (PET)/magnetic resonance (MR) acquisition on a fully integrated PET/MR scanner in patients and to compare the quality of PET images acquired with a PET/MR device with those acquired with a PET/computed tomography (CT) scanner. PATIENTS AND METHODS: Sequential PET/CT and PET/MR imaging was performed in 24 patients with neuroendocrine tumors using a single-injection/dual-imaging protocol. After intravenous injection of Ga-DOTATOC (mean, 120 MBq), PET/CT imaging including low-dose CT was performed at a mean time of 17 minutes post injection, and subsequently, PET/MR imaging including a Dixon sequence for attenuation correction was started at a mean time of 82 minutes post injection. The PET/CT and PET/MR images were analyzed visually using a 4-point scale for quality, coregistration, anatomical correlation, and lesion conspicuity. The standardized uptake value of background organs and focal lesions was measured and compared between the PET/CT and PET/MR acquisitions. RESULTS: Ga-DOTATOC PET acquired on the PET/MR delivered images with a good diagnostic quality (average visual rating PET/CT, 2.83; PET/MR, 2.08; P <; 0.01). The standardized uptake value of focal lesions did not differ between the PET/CT and PET/MR acquisitions (P >; 0.3) and correlated in a linear fashion (correlation coefficient ρ = 0.90). Lesion conspicuity was slightly, but significantly, higher on the PET/CT acquisitions (PET/CT, 2.71; PET/MR, 2.62; P = 0.01). Positron emission tomography/MR detected 153 of 157 lesions identified by PET/CT; however, there was no difference in sensitivity on a patient basis or organ system basis. Anatomical correlates for focal PET lesions could significantly more often be delineated using MR Dixon images compared with low-dose CT (average visual rating PET/CT, 1.78; PET/MR, 2.30; P <; 0.01). Coregistration of functional and morphological data was better on PET/MR compared with PET/CT, which, however, did not reach significance (average visual rating PET/CT, 2.17; PET/MR, 2.46; P = 0.10). CONCLUSIONS: Ga-DOTATOC PET/MR imaging is feasible in patients, with good image quality, and detectability of focal PET lesions was equivalent to PET/CT on a patient basis and organ system basis. Now, the clinical value of Ga-DOTATOC PET/MR with additional diagnostic MR protocols has to be evaluated against PET/CT with multiphase contrast-enhanced CT protocols in future studies.
KW - 68Ga-DOTATOC
KW - PET/CT
KW - PET/MR
KW - hybrid imaging
KW - neuroendocrine tumors
UR - http://www.scopus.com/inward/record.url?scp=84876296678&partnerID=8YFLogxK
U2 - 10.1097/RLI.0b013e31828234d0
DO - 10.1097/RLI.0b013e31828234d0
M3 - Article
C2 - 23385399
AN - SCOPUS:84876296678
SN - 0020-9996
VL - 48
SP - 263
EP - 272
JO - Investigative Radiology
JF - Investigative Radiology
IS - 5
ER -