Electron dosimetry based on the absorbed dose to water concept: A comparison of the AAPM TG-51 and DIN 6800-2 protocols

Oliver S. Dohm, Gunter Christ, Fridtjof Nüsslin, Edmund Schüle, Gregor Bruggmoser

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

The dosimetry protocols DIN 6800-2 and AAPM TG-51, both based on the absorbed dose to water concept, are compared in their theoretical background and in their application to electron dosimetry. The agreement and disagreement in correction factors and energy parameters used in both protocols will be shown and discussed. Measurements with three different types of ionization chambers were performed and evaluated according to both protocols. As a result the perturbation correction factor P60wallCo for the Roos chamber was determined to 1.024 ± 0.5%.

Original languageEnglish
Pages (from-to)2258-2264
Number of pages7
JournalMedical Physics
Volume28
Issue number11
DOIs
StatePublished - 2001
Externally publishedYes

Keywords

  • Absorbed dose to water
  • Electron dosimetry
  • Perturbation correction
  • Roos chamber