Abstract
In this article a new X-ray-free measurement method for postoperative 3D-position analysis of dental implants is proposed. The advantage of this new method is the possibility of precise 3D comparison to preoperative planning data without the use of X-rays. Standard methods for postoperative implant analysis are performed manually on the basis of orthopantomography or computer tomography scans. The proposed method uses a navigation system and a specially designed measurement device. Analysis of the measurement data shows a mean position error of 0.2 mm±0.1 mm and a mean depth error of 0.4 mm±0.3 mm. Thus, this method is suitable for postoperative analysis of the horizontal implant position and for comparison to preoperative planning.
Translated title of the contribution | A new X-ray-free measurement method for postoperative 3D-position analysis of navigated inserted implants |
---|---|
Original language | German |
Pages (from-to) | 234-242 |
Number of pages | 9 |
Journal | Biomedizinische Technik |
Volume | 52 |
Issue number | 3 |
DOIs | |
State | Published - 1 Jun 2007 |