Ein neues verfahren zur bildfreien kontrolle der postoperativen genauigkeit navigiert gesetzter implantate

Translated title of the contribution: A new X-ray-free measurement method for postoperative 3D-position analysis of navigated inserted implants

Sebastian Stopp, Tim Lüth

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

In this article a new X-ray-free measurement method for postoperative 3D-position analysis of dental implants is proposed. The advantage of this new method is the possibility of precise 3D comparison to preoperative planning data without the use of X-rays. Standard methods for postoperative implant analysis are performed manually on the basis of orthopantomography or computer tomography scans. The proposed method uses a navigation system and a specially designed measurement device. Analysis of the measurement data shows a mean position error of 0.2 mm±0.1 mm and a mean depth error of 0.4 mm±0.3 mm. Thus, this method is suitable for postoperative analysis of the horizontal implant position and for comparison to preoperative planning.

Translated title of the contributionA new X-ray-free measurement method for postoperative 3D-position analysis of navigated inserted implants
Original languageGerman
Pages (from-to)234-242
Number of pages9
JournalBiomedizinische Technik
Volume52
Issue number3
DOIs
StatePublished - 1 Jun 2007

Fingerprint

Dive into the research topics of 'A new X-ray-free measurement method for postoperative 3D-position analysis of navigated inserted implants'. Together they form a unique fingerprint.

Cite this