T1 - Effects of a dual-filter-based cerebral embolic protection device in transcatheter aortic valve replacement on cerebral oxygen saturation
T2 - A prospective pilot study
AU - Voss, Stephanie
AU - Ernst, Annick
AU - Erlebach, Magdalena
AU - Ruge, Hendrik
AU - Sideris, Konstantinos
AU - Bleiziffer, Sabine
AU - Voss, Bernhard
AU - Tassani-Prell, Peter
AU - Mayr, N. Patrick
N1 - Publisher Copyright:
© 2021 The Authors. Journal of Cardiac Surgery published by Wiley Periodicals LLC
PY - 2021/4
Y1 - 2021/4
N2 - Purpose: The Sentinel Cerebral Protection System (Sentinel-CPS) is increasingly used in transcatheter aortic valve replacement (TAVR). However, the impact of inserting the Sentinel-CPS inside the brain-supplying arteries on cerebral perfusion and oxygenation is unknown. Methods: Twenty patients undergoing transfemoral TAVR with (n = 10) and without (n = 10) cerebral embolic protection using the Sentinel-CPS were prospectively observed. All patients received conscious sedation and cerebral oxygen saturation (rSO2) was continuously measured with near-infrared spectroscopy (NIRS). The cumulative perioperative cerebral desaturation was calculated for each patient by multiplying rSO2 below an individualized desaturation threshold by time. In addition, rSO2 values at the time of Sentinel-CPS insertion, filter positioning, and device retraction were analyzed. Results: There was no significant difference in cumulative cerebral desaturation in patients with Sentinel-CPS (median [IQR]) (0 [0/81] s%) and without (median [IQR]) (0 [0/23] s%), p =.762. A total of 6 patients (33.3%) experienced a perioperative decrease in rSO2 below the individualized desaturation threshold (n = 3 with Sentinel-CPS, n = 3 without Sentinel-CPS; p = 1.000). Cerebral desaturation was detected during valve deployment (n = 5) and after postdilatation (n = 1). No desaturation events occurred during Sentinel-CPS insertion, filter positioning, or retraction. Conclusion: Our pilot study revealed no difference in cumulative perioperative cerebral desaturation between TAVR with and without Sentinel-CPS. Catheter- and filter-based manipulations in the brain-supplying arteries for Sentinel-CPS application were not associated with a decrease of cerebral perfusion and oxygenation.
AB - Purpose: The Sentinel Cerebral Protection System (Sentinel-CPS) is increasingly used in transcatheter aortic valve replacement (TAVR). However, the impact of inserting the Sentinel-CPS inside the brain-supplying arteries on cerebral perfusion and oxygenation is unknown. Methods: Twenty patients undergoing transfemoral TAVR with (n = 10) and without (n = 10) cerebral embolic protection using the Sentinel-CPS were prospectively observed. All patients received conscious sedation and cerebral oxygen saturation (rSO2) was continuously measured with near-infrared spectroscopy (NIRS). The cumulative perioperative cerebral desaturation was calculated for each patient by multiplying rSO2 below an individualized desaturation threshold by time. In addition, rSO2 values at the time of Sentinel-CPS insertion, filter positioning, and device retraction were analyzed. Results: There was no significant difference in cumulative cerebral desaturation in patients with Sentinel-CPS (median [IQR]) (0 [0/81] s%) and without (median [IQR]) (0 [0/23] s%), p =.762. A total of 6 patients (33.3%) experienced a perioperative decrease in rSO2 below the individualized desaturation threshold (n = 3 with Sentinel-CPS, n = 3 without Sentinel-CPS; p = 1.000). Cerebral desaturation was detected during valve deployment (n = 5) and after postdilatation (n = 1). No desaturation events occurred during Sentinel-CPS insertion, filter positioning, or retraction. Conclusion: Our pilot study revealed no difference in cumulative perioperative cerebral desaturation between TAVR with and without Sentinel-CPS. Catheter- and filter-based manipulations in the brain-supplying arteries for Sentinel-CPS application were not associated with a decrease of cerebral perfusion and oxygenation.
KW - TAVR
KW - cerebral embolic protection
KW - cerebral oximetry
KW - dual-filter system