Comparison of Acute Thrombogenicity for Metallic and Polymeric Bioabsorbable Scaffolds: Magmaris Versus Absorb in a Porcine Arteriovenous Shunt Model

Ron Waksman, Michael J. Lipinski, Eduardo Acampado, Qi Cheng, Lila Adams, Sho Torii, Jiaxiang Gai, Rebecca Torguson, David M. Hellinga, Peter C. Westman, Michael Joner, Philine Zumstein, Frank D. Kolodgie, Renu Virmani

Research output: Contribution to journalArticlepeer-review

92 Scopus citations

Abstract

METHODS AND RESULTS: An ex vivo porcine carotid jugular arteriovenous shunt was established and connected to SYLGARD tubing containing the Magmaris, Absorb, and Orsiro scaffolds/stents and allowed to run in the shunt for a maximum of 1 hour. Twelve shunts (2 shunt runs per pig) were run comparing the 3 scaffolds in alternating order. Nested generalized linear mixed models were used to compare variables between scaffold groups while adjusting for variability between shunt runs. Confocal fluorescent microscopy costaining CD61/CD42b demonstrated that both Magmaris (3.0%) and Orsiro (4.6%) had less platelet coverage of the total scaffold compared with Absorb (21.8%). Scanning electron microscopy demonstrated significantly less thrombus deposition to Magmaris as a percentage of the total scaffold compared with Absorb (5.0% versus 16.1%, P=0.02). Magmaris had significantly less PM-1-positive neutrophil and CD14-positive monocyte adherence compared with both Orsiro and Absorb. Orsiro had significantly less monocyte deposition compared with Absorb.

CONCLUSIONS: Despite a similar scaffold strut thickness, the Magmaris sirolimus-eluting bioabsorbable magnesium scaffold was significantly less thrombogenic compared with the Absorb bioresorbable vascular scaffold in an ex vivo porcine arteriovenous shunt model. Further studies are needed to determine whether the reduced thrombogenicity of Magmaris will result in reductions in major cardiovascular events.

BACKGROUND: A comparison in acute thrombogenicity between the Magmaris sirolimus-eluting bioabsorbable magnesium scaffold and the Absorb bioresorbable vascular scaffold has not been performed. This study assessed acute thrombogenicity of Magmaris compared with Absorb and the Orsiro hybrid drug-eluting stent in a porcine arteriovenous shunt model.

Original languageEnglish
JournalCirculation: Cardiovascular Interventions
Volume10
Issue number8
DOIs
StatePublished - 1 Aug 2017
Externally publishedYes

Keywords

  • acute thrombogenicity
  • bioresorbable scaffold
  • drug-eluting stents
  • monocytes
  • neutrophils