TY - JOUR
T1 - Comparison of a Pure Plug-Based Versus a Primary Suture-Based Vascular Closure Device Strategy for Transfemoral Transcatheter Aortic Valve Replacement
T2 - The CHOICE-CLOSURE Randomized Clinical Trial
AU - Abdel-Wahab, Mohamed
AU - Hartung, Philipp
AU - Dumpies, Oliver
AU - Obradovic, Danilo
AU - Wilde, Johannes
AU - Majunke, Nicolas
AU - Boekstegers, Peter
AU - Müller, Ralf
AU - Seyfarth, Melchior
AU - Vorpahl, Marc
AU - Kiefer, Philipp
AU - Noack, Thilo
AU - Leontyev, Sergey
AU - Sandri, Marcus
AU - Rotta Detto Loria, Johannes
AU - Kitamura, Mitsunobu
AU - Borger, Michael Andrew
AU - Funkat, Anne Kathrin
AU - Hohenstein, Sven
AU - Desch, Steffen
AU - Holzhey, David
AU - Thiele, Holger
N1 - Publisher Copyright:
© 2022 Lippincott Williams and Wilkins. All rights reserved.
PY - 2022/1/18
Y1 - 2022/1/18
N2 - Background: Transcatheter aortic valve replacement is an established treatment option for patients with severe symptomatic aortic stenosis and is most commonly performed through the transfemoral access route. Percutaneous access site closure can be achieved using dedicated plug-based or suture-based vascular closure device (VCD) strategies, but randomized comparative studies are scarce. Methods: The CHOICE-CLOSURE trial (Randomized Comparison of Catheter-based Strategies for Interventional Access Site Closure during Transfemoral Transcatheter Aortic Valve Implantation) is an investigator-initiated, multicenter study, in which patients undergoing transfemoral transcatheter aortic valve replacement were randomly assigned to vascular access site closure using either a pure plug-based technique (MANTA, Teleflex) with no additional VCDs or a primary suture-based technique (ProGlide, Abbott Vascular) potentially complemented by a small plug. The primary end point consisted of access site- or access-related major and minor vascular complications during index hospitalization, defined according to the Valve Academic Research Consortium-2 criteria. Secondary end points included the rate of access site- or access-related bleeding, VCD failure, and time to hemostasis. Results: A total of 516 patients were included and randomly assigned. The mean age of the study population was 80.5±6.1 years, 55.4% were male, 7.6% of patients had peripheral vascular disease, and the mean Society of Thoracic Surgeons score was 4.1±2.9%. The primary end point occurred in 19.4% (50/258) of the pure plug-based group and 12.0% (31/258) of the primary suture-based group (relative risk, 1.61 [95% CI, 1.07-2.44], P=0.029). Access site- or access-related bleeding occurred in 11.6% versus 7.4% (relative risk, 1.58 [95%CI: 0.91-2.73], P=0.133) and device failure in 4.7% versus 5.4% (relative risk, 0.86, [95% CI, 0.40-1.82], P=0.841) in the respective groups. Time to hemostasis was significantly shorter in the pure plug-based group (80 [32-180] versus 240 [174-316] seconds, P<0.001). Conclusions: Among patients treated with transfemoral transcatheter aortic valve replacement, a pure plug-based vascular closure technique using the MANTA VCD is associated with a higher rate of access site- or access-related vascular complications but a shorter time to hemostasis compared with a primary suture-based technique using the ProGlide VCD.
AB - Background: Transcatheter aortic valve replacement is an established treatment option for patients with severe symptomatic aortic stenosis and is most commonly performed through the transfemoral access route. Percutaneous access site closure can be achieved using dedicated plug-based or suture-based vascular closure device (VCD) strategies, but randomized comparative studies are scarce. Methods: The CHOICE-CLOSURE trial (Randomized Comparison of Catheter-based Strategies for Interventional Access Site Closure during Transfemoral Transcatheter Aortic Valve Implantation) is an investigator-initiated, multicenter study, in which patients undergoing transfemoral transcatheter aortic valve replacement were randomly assigned to vascular access site closure using either a pure plug-based technique (MANTA, Teleflex) with no additional VCDs or a primary suture-based technique (ProGlide, Abbott Vascular) potentially complemented by a small plug. The primary end point consisted of access site- or access-related major and minor vascular complications during index hospitalization, defined according to the Valve Academic Research Consortium-2 criteria. Secondary end points included the rate of access site- or access-related bleeding, VCD failure, and time to hemostasis. Results: A total of 516 patients were included and randomly assigned. The mean age of the study population was 80.5±6.1 years, 55.4% were male, 7.6% of patients had peripheral vascular disease, and the mean Society of Thoracic Surgeons score was 4.1±2.9%. The primary end point occurred in 19.4% (50/258) of the pure plug-based group and 12.0% (31/258) of the primary suture-based group (relative risk, 1.61 [95% CI, 1.07-2.44], P=0.029). Access site- or access-related bleeding occurred in 11.6% versus 7.4% (relative risk, 1.58 [95%CI: 0.91-2.73], P=0.133) and device failure in 4.7% versus 5.4% (relative risk, 0.86, [95% CI, 0.40-1.82], P=0.841) in the respective groups. Time to hemostasis was significantly shorter in the pure plug-based group (80 [32-180] versus 240 [174-316] seconds, P<0.001). Conclusions: Among patients treated with transfemoral transcatheter aortic valve replacement, a pure plug-based vascular closure technique using the MANTA VCD is associated with a higher rate of access site- or access-related vascular complications but a shorter time to hemostasis compared with a primary suture-based technique using the ProGlide VCD.
KW - randomized control trial
KW - sutures
KW - transcatheter aortic valve replacement
KW - vascular closure devices
UR - http://www.scopus.com/inward/record.url?scp=85121664840&partnerID=8YFLogxK
U2 - 10.1161/CIRCULATIONAHA.121.057856
DO - 10.1161/CIRCULATIONAHA.121.057856
M3 - Article
C2 - 34738828
AN - SCOPUS:85121664840
SN - 0009-7322
VL - 145
SP - 170
EP - 183
JO - Circulation
JF - Circulation
IS - 3
ER -