TY - JOUR
T1 - Comparison between balanced steady-state free precession and standard spoiled gradient echo magnetization transfer ratio imaging in multiple sclerosis
T2 - Methodical and clinical considerations
AU - Amann, Michael
AU - Sprenger, Till
AU - Naegelin, Yvonne
AU - Reinhardt, Julia
AU - Kuster, Pascal
AU - Hirsch, Jochen G.
AU - Kappos, Ludwig
AU - Radue, Ernst Wilhelm
AU - Stippich, Christoph
AU - Bieri, Oliver
N1 - Publisher Copyright:
© 2014 Elsevier Inc.
PY - 2015/3/1
Y1 - 2015/3/1
N2 - Different pathological processes like demyelination and axonal loss can alter the magnetisation transfer ratio (MTR) in brain tissue. The standard method to measure this effect is to scan the respective tissue twice, one with and one without a specific saturation pulse. A major drawback of this technique based on spoiled gradient echo (GRE) sequences relates to its long acquisition time due to the saturation pulses. Recently, an alternative concept for MT imaging based on balanced steady state free precession (bSSFP) has been proposed. Modification of the duration of the radiofrequency pulses for imaging allows scanning MT sensitive and non-sensitive images. The steady-state character of bSSFP with high intrinsic signal-to-noise ratio (SNR) allows three-dimensional (3D) whole brain MTR at high spatial resolution within short and thus clinically feasible acquisition times. In the present study, both bSSFP-MT and 2D GRE-MT imaging were used in a cohort of 31 patients with multiple sclerosis (MS) to characterize different normal appearing (NA) and pathological brain structures. Under the constraint of identical SNR and scan time, a 3.4 times higher voxel size could be achieved with bSSFP. This increased resolution allowed a more accurate delineation of the different brain structures, especially of cortex, hippocampus and MS lesions. In a multiple linear regression model, we found an association between MTR of cortical lesions and a clinical measure of disability (r. = -. 0.407, p. =. 0.035) in the bSSFP dataset only. The different relaxation weighting of the base images (T2/T1 in bSSFP, proton density in GRE) had no effects besides a larger spreading of the MTR values of the different NA structures. This was demonstrated by the nearly perfect linearity between the NA matter MTR of both techniques as well as in the absolute MTR differences between NA matter and the respective lesions.
AB - Different pathological processes like demyelination and axonal loss can alter the magnetisation transfer ratio (MTR) in brain tissue. The standard method to measure this effect is to scan the respective tissue twice, one with and one without a specific saturation pulse. A major drawback of this technique based on spoiled gradient echo (GRE) sequences relates to its long acquisition time due to the saturation pulses. Recently, an alternative concept for MT imaging based on balanced steady state free precession (bSSFP) has been proposed. Modification of the duration of the radiofrequency pulses for imaging allows scanning MT sensitive and non-sensitive images. The steady-state character of bSSFP with high intrinsic signal-to-noise ratio (SNR) allows three-dimensional (3D) whole brain MTR at high spatial resolution within short and thus clinically feasible acquisition times. In the present study, both bSSFP-MT and 2D GRE-MT imaging were used in a cohort of 31 patients with multiple sclerosis (MS) to characterize different normal appearing (NA) and pathological brain structures. Under the constraint of identical SNR and scan time, a 3.4 times higher voxel size could be achieved with bSSFP. This increased resolution allowed a more accurate delineation of the different brain structures, especially of cortex, hippocampus and MS lesions. In a multiple linear regression model, we found an association between MTR of cortical lesions and a clinical measure of disability (r. = -. 0.407, p. =. 0.035) in the bSSFP dataset only. The different relaxation weighting of the base images (T2/T1 in bSSFP, proton density in GRE) had no effects besides a larger spreading of the MTR values of the different NA structures. This was demonstrated by the nearly perfect linearity between the NA matter MTR of both techniques as well as in the absolute MTR differences between NA matter and the respective lesions.
KW - Balanced SSFP
KW - Lesions
KW - Magnetization transfer ratio
KW - Multiple sclerosis
KW - Normal appearing matter
UR - http://www.scopus.com/inward/record.url?scp=84922617255&partnerID=8YFLogxK
U2 - 10.1016/j.neuroimage.2014.12.045
DO - 10.1016/j.neuroimage.2014.12.045
M3 - Article
C2 - 25536494
AN - SCOPUS:84922617255
SN - 1053-8119
VL - 108
SP - 87
EP - 94
JO - NeuroImage
JF - NeuroImage
ER -