Comparing physiologically relevant corrosion performances of Mg AZ31 alloy protected by ALD and sputter coated TiO2

Mirco Peron, Abdulla Bin Afif, Anup Dadlani, Filippo Berto, Jan Torgersen

Research output: Contribution to journalArticlepeer-review

35 Scopus citations

Abstract

The utilization of Mg alloys for biomedical applications is so far underexplored due to the accelerated corrosion hampering patient recovery post implantation. Here, we explore the effectiveness of corrosion reduction of an AZ31 alloy in Simulated Body Fluid when coated with a 40 nm sputtered TiO2 layer and compare it to a similar coating made by Atomic Layer Deposition (ALD). Potentiodynamic polarization and hydrogen evolution experiments were performed on coated samples having different surface roughness and 3D topologies. Interestingly, ALD layers reduce corrosion current density by 94% on Ra = 118.6 ± 5.1 nm and 93% on Ra = 4794.3 ± 49.4 nm, whereas sputtered only by 84% on Ra = 118.6 ± 5.1 nm and 60% on Ra = 4794.3 ± 49.4 nm. Particularly on 3D aspects, the ALD coatings are superior, where a scaffold of 85% porosity with 1 mm pore sizes released 68% lower hydrogen compared to the sputtered counterparts. We relate these observations to the higher surface integrity, adhesion strength and lower line-of-sight restrictions of ALD compared to sputter deposition. The results can be interesting for researchers and practitioners aiming to make Mg alloys more commonplace as temporary metallic implant materials.

Original languageEnglish
Article number125922
JournalSurface and Coatings Technology
Volume395
DOIs
StatePublished - 15 Aug 2020
Externally publishedYes

Keywords

  • Atomic Layer Deposition (ALD)
  • Biocompatible coatings
  • Corrosion resistance
  • Magnesium alloys
  • Sputter
  • Temporary metallic implants

Fingerprint

Dive into the research topics of 'Comparing physiologically relevant corrosion performances of Mg AZ31 alloy protected by ALD and sputter coated TiO2'. Together they form a unique fingerprint.

Cite this