Skip to main navigation Skip to search Skip to main content

Clinically relevant lot-to-lot reagent difference in a commercial immunoturbidimetric assay for glycated hemoglobin A1c

  • Technical University of Munich

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Objectives: Hemoglobin A1c (HbA1c) is employed for diagnosis and therapy monitoring of diabetes mellitus. The effect of a change of reagent lot on the measured values of a commercial immunoturbidimetric HbA1c assay (A1C3) was investigated. Design and methods: Comparison measurements of A1C3 and an automated affinity chromatography method (VIIT) were performed in 15 samples for the initial and in 20 samples for the subsequent A1C3 lot. The results of 27 and 19 measurements of a normal and of 28 and 20 of a pathological control (before and after the switch of the A1C3 reagent lot, respectively) were evaluated. Finally, the results of 6463 patient samples that had been measured with the initial and 434 that had been measured with the subsequent A1C3 lot were investigated. Results: VIIT yielded significantly higher results than the initial A1C3 lot (bias: 0.41% HbA1c, 4.5. mmol/mol) but agreed well with the subsequent lot (bias: - 0.01% HbA1c, - 0.1. mmol/mol). Changing to the subsequent reagent lot resulted in significant increases of the mean of the normal control of 0.316% HbA1c (3.5. mmol/mol) and of the pathological control of 0.749% HbA1c (8.2. mmol/mol). The median of patient samples measured with the subsequent lot was significantly higher by 0.40% HbA1c (4.4. mmol/mol). Conclusions: The subsequent A1C3 reagent lot yields significantly higher measurement results than the initial by approximately 0.5% HbA1c (5.5. mmol/mol). This difference is considered as clinically relevant. A combined effort of manufacturers and notified bodies is necessary to minimize lot-to-lot variation.

Original languageEnglish
Pages (from-to)1167-1170
Number of pages4
JournalClinical Biochemistry
Volume48
Issue number16-17
DOIs
StatePublished - 28 May 2015

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • HbA1c
  • Immunoassay
  • Lot variability

Fingerprint

Dive into the research topics of 'Clinically relevant lot-to-lot reagent difference in a commercial immunoturbidimetric assay for glycated hemoglobin A1c'. Together they form a unique fingerprint.

Cite this