TY - JOUR
T1 - Cement-Augmented Carbon Fiber–Reinforced Pedicle Screw Instrumentation for Spinal Metastases
T2 - Safety and Efficacy
AU - Wagner, Arthur
AU - Haag, Elena
AU - Joerger, Ann Kathrin
AU - Gempt, Jens
AU - Krieg, Sandro M.
AU - Wostrack, Maria
AU - Meyer, Bernhard
N1 - Publisher Copyright:
© 2021 Elsevier Inc.
PY - 2021/10
Y1 - 2021/10
N2 - Objective: To investigate the complication rates and long-term implant failure rates in a monocentric study of a consecutive cohort of patients with thoracolumbar spinal metastases after posterior instrumentation with a fenestrated carbon fiber–reinforced poly-ether-ether-ketone (CFRP) pedicle screw system. Methods: We retrospectively reviewed demographics, Karnofsky Performance Status Scale scores, complications, and implant failure rates. Results: Between June 2016 and November 2019, 51 consecutive patients underwent cement-augmented CFRP pedicle screw instrumentation at our institution. Mean age was 68 years (standard deviation 10.5), the median preoperative Karnofsky Performance Status Scale of 80 increased to 90 postoperatively (P = 0.471). Most common primary entities were breast (25.5%), lung (15.7%), and prostate (13.7%) cancers. Of 428 placed screws, 293 (68.5%) were augmented with polymethylmethacrylate, a mean 6 per patient (standard deviation ±2). Screws were inserted via a minimally invasive system technique in 54.9% of cases. In total, 11.8% of patients had immediate postoperative sequelae related to the cement. Pulmonary cement embolisms were noted in 3 patients, 2 had paravertebral extravasation, and 1 had an embolism into a segmental artery. Of these 6, 2 patients with pulmonary embolisms reported related symptoms. Follow-up was available for 80.4%. After a mean 9.8 months, screw loosening was noted in 11.8% of cases on computed tomography, although it was asymptomatic in all but 1 patient. Screw pull-out did not occur. Neither cement-related (P = 0.353) nor general complication rates (P = 0.507) differed significantly between open and minimally invasive system techniques. Conclusions: Percutaneous cement-augmented CFRP pedicle screw instrumentation facilitates artifact-reduced postoperative imaging, while maintaining a risk profile and implant failure rates comparable to conventional metallic instrumentation.
AB - Objective: To investigate the complication rates and long-term implant failure rates in a monocentric study of a consecutive cohort of patients with thoracolumbar spinal metastases after posterior instrumentation with a fenestrated carbon fiber–reinforced poly-ether-ether-ketone (CFRP) pedicle screw system. Methods: We retrospectively reviewed demographics, Karnofsky Performance Status Scale scores, complications, and implant failure rates. Results: Between June 2016 and November 2019, 51 consecutive patients underwent cement-augmented CFRP pedicle screw instrumentation at our institution. Mean age was 68 years (standard deviation 10.5), the median preoperative Karnofsky Performance Status Scale of 80 increased to 90 postoperatively (P = 0.471). Most common primary entities were breast (25.5%), lung (15.7%), and prostate (13.7%) cancers. Of 428 placed screws, 293 (68.5%) were augmented with polymethylmethacrylate, a mean 6 per patient (standard deviation ±2). Screws were inserted via a minimally invasive system technique in 54.9% of cases. In total, 11.8% of patients had immediate postoperative sequelae related to the cement. Pulmonary cement embolisms were noted in 3 patients, 2 had paravertebral extravasation, and 1 had an embolism into a segmental artery. Of these 6, 2 patients with pulmonary embolisms reported related symptoms. Follow-up was available for 80.4%. After a mean 9.8 months, screw loosening was noted in 11.8% of cases on computed tomography, although it was asymptomatic in all but 1 patient. Screw pull-out did not occur. Neither cement-related (P = 0.353) nor general complication rates (P = 0.507) differed significantly between open and minimally invasive system techniques. Conclusions: Percutaneous cement-augmented CFRP pedicle screw instrumentation facilitates artifact-reduced postoperative imaging, while maintaining a risk profile and implant failure rates comparable to conventional metallic instrumentation.
KW - Carbon
KW - Dement augmentation
KW - Instrumentation
KW - PEEK
KW - Spinal metastasis
UR - http://www.scopus.com/inward/record.url?scp=85112757753&partnerID=8YFLogxK
U2 - 10.1016/j.wneu.2021.07.092
DO - 10.1016/j.wneu.2021.07.092
M3 - Article
C2 - 34339894
AN - SCOPUS:85112757753
SN - 1878-8750
VL - 154
SP - e536-e546
JO - World Neurosurgery
JF - World Neurosurgery
ER -