Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900

  • Hendrik Ruge
  • , Hector A. Alvarez-Covarrubias
  • , Oliver Deutsch
  • , Zahra Alalawi
  • , Keti Vitanova
  • , Rüdiger Lange

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Background: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. Methods: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF. Results: High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect. Conclusion: Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics.

Original languageEnglish
Article number859088
JournalFrontiers in Cardiovascular Medicine
Volume9
DOIs
StatePublished - 2 Jun 2022

Keywords

  • balloon rupture
  • bioprosthetic valve fracturing
  • in-vitro
  • transprosthetic gradient
  • valve-in-valve transcatheter aortic valve replacement

Fingerprint

Dive into the research topics of 'Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900'. Together they form a unique fingerprint.

Cite this