TY - JOUR
T1 - ASD and PFO closure with the Solysafe Septal Occluder - Results of a prospective multicenter pilot study
AU - Ewert, Peter
AU - Söderberg, Bjoern
AU - Dähnert, Ingo
AU - Hess, Otto
AU - Schuler, Gerhard
AU - Bussmann, Christiane
AU - Bernhard, Jerome
AU - Sick, Peter
PY - 2008/2/5
Y1 - 2008/2/5
N2 - Objective: We report the results of a prospective multicenter pilot study performed in Germany, Sweden, and Switzerland with a new self-centering device for transcatheter closure of an atrial septal defect (ASD) or a patent foramen ovale (PFO) called the Solysafe Septal Occluder. Interventions: The device was successfully implanted in 44 patients. In 15 patients with a median age of 40 years (range 6-76 years), a Solysafe device was successfully implanted in an ASD. The median size of the stretched defects was 17 mm (range 10-21 mm). Three 15-mm devices, eight 20-mm devices, and four 25-mm devices were used. Procedure time ranged from 40 to 107 min (median 66 min) and fluoroscopic time from 5.3 to 17.5 min (median 12 min). In 29 patients with a median age of 47 years (range 15-78 years), a Solysafe device was implanted in a PFO. The procedure time ranged from 21 to 155 min (median 51 min) and fluoroscopic time from 3.1 to 31.3 min (median 7.6 min). Results: At discharge, 1 of the 29 patients (3%) had a small shunt. No patient in either the ASD or the PFO group had any major complication. Six months after implantation, the overall closure rate with the Solysafe septal occluder in both groups was 100% (44/44). Conclusion: With the self-centering Solysafe Septal Occluder, PFOs, and ASDs with a stretched diameter of up to 21 mm can be effectively closed with very high occlusion rates.
AB - Objective: We report the results of a prospective multicenter pilot study performed in Germany, Sweden, and Switzerland with a new self-centering device for transcatheter closure of an atrial septal defect (ASD) or a patent foramen ovale (PFO) called the Solysafe Septal Occluder. Interventions: The device was successfully implanted in 44 patients. In 15 patients with a median age of 40 years (range 6-76 years), a Solysafe device was successfully implanted in an ASD. The median size of the stretched defects was 17 mm (range 10-21 mm). Three 15-mm devices, eight 20-mm devices, and four 25-mm devices were used. Procedure time ranged from 40 to 107 min (median 66 min) and fluoroscopic time from 5.3 to 17.5 min (median 12 min). In 29 patients with a median age of 47 years (range 15-78 years), a Solysafe device was implanted in a PFO. The procedure time ranged from 21 to 155 min (median 51 min) and fluoroscopic time from 3.1 to 31.3 min (median 7.6 min). Results: At discharge, 1 of the 29 patients (3%) had a small shunt. No patient in either the ASD or the PFO group had any major complication. Six months after implantation, the overall closure rate with the Solysafe septal occluder in both groups was 100% (44/44). Conclusion: With the self-centering Solysafe Septal Occluder, PFOs, and ASDs with a stretched diameter of up to 21 mm can be effectively closed with very high occlusion rates.
KW - Atrial septal defect
KW - Persistent foramen ovale
KW - Self centering device
KW - Transcatheter therapy
UR - http://www.scopus.com/inward/record.url?scp=41149130307&partnerID=8YFLogxK
U2 - 10.1002/ccd.21360
DO - 10.1002/ccd.21360
M3 - Article
C2 - 18288739
AN - SCOPUS:41149130307
SN - 1522-1946
VL - 71
SP - 398
EP - 402
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
IS - 3
ER -