ASD and PFO closure with the Solysafe Septal Occluder - Results of a prospective multicenter pilot study

Peter Ewert, Bjoern Söderberg, Ingo Dähnert, Otto Hess, Gerhard Schuler, Christiane Bussmann, Jerome Bernhard, Peter Sick

Research output: Contribution to journalArticlepeer-review

26 Scopus citations

Abstract

Objective: We report the results of a prospective multicenter pilot study performed in Germany, Sweden, and Switzerland with a new self-centering device for transcatheter closure of an atrial septal defect (ASD) or a patent foramen ovale (PFO) called the Solysafe Septal Occluder. Interventions: The device was successfully implanted in 44 patients. In 15 patients with a median age of 40 years (range 6-76 years), a Solysafe device was successfully implanted in an ASD. The median size of the stretched defects was 17 mm (range 10-21 mm). Three 15-mm devices, eight 20-mm devices, and four 25-mm devices were used. Procedure time ranged from 40 to 107 min (median 66 min) and fluoroscopic time from 5.3 to 17.5 min (median 12 min). In 29 patients with a median age of 47 years (range 15-78 years), a Solysafe device was implanted in a PFO. The procedure time ranged from 21 to 155 min (median 51 min) and fluoroscopic time from 3.1 to 31.3 min (median 7.6 min). Results: At discharge, 1 of the 29 patients (3%) had a small shunt. No patient in either the ASD or the PFO group had any major complication. Six months after implantation, the overall closure rate with the Solysafe septal occluder in both groups was 100% (44/44). Conclusion: With the self-centering Solysafe Septal Occluder, PFOs, and ASDs with a stretched diameter of up to 21 mm can be effectively closed with very high occlusion rates.

Original languageEnglish
Pages (from-to)398-402
Number of pages5
JournalCatheterization and Cardiovascular Interventions
Volume71
Issue number3
DOIs
StatePublished - 5 Feb 2008
Externally publishedYes

Keywords

  • Atrial septal defect
  • Persistent foramen ovale
  • Self centering device
  • Transcatheter therapy

Fingerprint

Dive into the research topics of 'ASD and PFO closure with the Solysafe Septal Occluder - Results of a prospective multicenter pilot study'. Together they form a unique fingerprint.