T1 - Application of right atrial contiguous linear lesions
T2 - An in vivo efficacy validation of multipolar ablation catheters in an animal model
AU - Schneider, Michael A.E.
AU - Ndrepepa, Gjin
AU - Vallant, Andre
AU - Gayk, Ute
AU - Richter, Thomas
AU - Henke, Julia
AU - Zrenner, Bernhard
AU - Karch, Martin R.
AU - Erhardt, Wolfgang
AU - Schömig, Albert
AU - Schmitt, Claus
PY - 2002/10/1
Y1 - 2002/10/1
N2 - Catheter ablation of atrial flutter and catheter Maze procedures require the creation of linear lesions. The efficacy of different multipolar catheters to create linear contiguous transmural lesions was studied in a sheep model. In 19 sheep a multipolar ablation catheter was inserted into the right atrium. In nine sheep a steerable 7 Fr catheter (C1) with six 6-mm electrodes and in five sheep a steerable 7 Fr catheter (C2) with four 5-mm electrodes were placed. In five sheep a 3.7 Fr catheter (C3) with eight electrodes of 6-mm length was deployed (steerable sheath). A total of 72 linear lesions were created and evaluated. Catheter types C1, C2, and C3 produced continuous lesions with at least two adjacent electrodes in 45%, 79%, and 87% of the lesions. The mean lesion length achieved by C3 was longer compared to C1 and C2 (27 ± 14 vs 10 ± 5 and 11 ± 6 mm; P < 0.05). The ability to produce contiguous lesions by all available electrodes was low: C1, 5%; C2, 5%; and C3, 6%. C3 was most effective in exerting transmural lesions (93% vs C1 75% and C2 57%; P < 0.0001). Microscopic endocardial fibrinous adhesions and macroscopic mild electrode carbonizations were caused by all catheter types. In conclusion, (1) all three catheter types do not create contiguous lesions along all electrodes. Gaps of viable tissue remain in most instances; (2) lesion depths and transmurality varies with different catheters; and (3) potentially hazardous thrombotic material was observed during radiofrequency ablation with all three catheters.
AB - Catheter ablation of atrial flutter and catheter Maze procedures require the creation of linear lesions. The efficacy of different multipolar catheters to create linear contiguous transmural lesions was studied in a sheep model. In 19 sheep a multipolar ablation catheter was inserted into the right atrium. In nine sheep a steerable 7 Fr catheter (C1) with six 6-mm electrodes and in five sheep a steerable 7 Fr catheter (C2) with four 5-mm electrodes were placed. In five sheep a 3.7 Fr catheter (C3) with eight electrodes of 6-mm length was deployed (steerable sheath). A total of 72 linear lesions were created and evaluated. Catheter types C1, C2, and C3 produced continuous lesions with at least two adjacent electrodes in 45%, 79%, and 87% of the lesions. The mean lesion length achieved by C3 was longer compared to C1 and C2 (27 ± 14 vs 10 ± 5 and 11 ± 6 mm; P < 0.05). The ability to produce contiguous lesions by all available electrodes was low: C1, 5%; C2, 5%; and C3, 6%. C3 was most effective in exerting transmural lesions (93% vs C1 75% and C2 57%; P < 0.0001). Microscopic endocardial fibrinous adhesions and macroscopic mild electrode carbonizations were caused by all catheter types. In conclusion, (1) all three catheter types do not create contiguous lesions along all electrodes. Gaps of viable tissue remain in most instances; (2) lesion depths and transmurality varies with different catheters; and (3) potentially hazardous thrombotic material was observed during radiofrequency ablation with all three catheters.
KW - Atrial fibrillation
KW - Atrium
KW - Linear lesion
KW - Radiofrequency ablation
UR - http://www.scopus.com/inward/record.url?scp=0036802420&partnerID=8YFLogxK
U2 - 10.1046/j.1460-9592.2002.01459.x
DO - 10.1046/j.1460-9592.2002.01459.x
M3 - Article
C2 - 12418744
AN - SCOPUS:0036802420
SN - 0147-8389
VL - 25
SP - 1459
EP - 1466
JO - Pacing and Clinical Electrophysiology
JF - Pacing and Clinical Electrophysiology
IS - 10
ER -