TY - JOUR
T1 - Aortic diameter assessment by cardiovascular magnetic resonance
T2 - Do we really need contrast enhanced images?
AU - Ferrari, Irene
AU - Shehu, Nerejda
AU - Mkrtchyan, Naira
AU - Latus, Heiner
AU - Stern, Heiko
AU - Martinoff, Stefan
AU - Ewert, Peter
AU - Meierhofer, Christian
N1 - Publisher Copyright:
© Cardiovascular Diagnosis and Therapy. All rights reserved.
PY - 2021/12
Y1 - 2021/12
N2 - Background: Cardiovascular magnetic resonance (CMR) is widely used for aortic diameter assessment but there is no consensus on the sequence or cardiac cycle phase in which the measures should be taken. The most used sequence is contrast-enhanced-magnetic-resonance-angiography (angiography), usually non-ECG-triggered. An alternative is a navigated 3D-whole-heart-steady-state-free-precession sequence which is contrast-free and breath- and ECG-gated (mostly diastolic gating), producing very sharp anatomical rendering. Nonetheless, its routine use has not yet spread. Our aim was evaluating aortic diameters by a systolic-gated 3D and put additional effort in the validation of diastolic-gated 3D as alternative to angiography. Methods: We retrospectively analysed 30 patients scheduled for routine Angiography. We measured the aorta at 9 standard positions by three different sequences (angiography, 3D-diastole and 3D-systole) and compared the diameters obtained by calculating the differences and by paired t-test analysis. Results: Diameters by 3D-systole were larger than by 3D-diastole and angiography (P<0.01). In the ascending aorta we found the maximal differences between systole and diastole and between systole and angiography which were 1.7± SD 1.02 mm and 1.5± SD 1.07 mm respectively. There was no significant difference between diastolic and angiography measurements (mean difference 0.2± SD 0.16 mm, P not significant). Conclusions: Our results support the use of navigated 3D-whole-heart CMR to evaluate aortic diameters. Systolic-gated 3D produces larger diameter, especially in the ascending aorta.
AB - Background: Cardiovascular magnetic resonance (CMR) is widely used for aortic diameter assessment but there is no consensus on the sequence or cardiac cycle phase in which the measures should be taken. The most used sequence is contrast-enhanced-magnetic-resonance-angiography (angiography), usually non-ECG-triggered. An alternative is a navigated 3D-whole-heart-steady-state-free-precession sequence which is contrast-free and breath- and ECG-gated (mostly diastolic gating), producing very sharp anatomical rendering. Nonetheless, its routine use has not yet spread. Our aim was evaluating aortic diameters by a systolic-gated 3D and put additional effort in the validation of diastolic-gated 3D as alternative to angiography. Methods: We retrospectively analysed 30 patients scheduled for routine Angiography. We measured the aorta at 9 standard positions by three different sequences (angiography, 3D-diastole and 3D-systole) and compared the diameters obtained by calculating the differences and by paired t-test analysis. Results: Diameters by 3D-systole were larger than by 3D-diastole and angiography (P<0.01). In the ascending aorta we found the maximal differences between systole and diastole and between systole and angiography which were 1.7± SD 1.02 mm and 1.5± SD 1.07 mm respectively. There was no significant difference between diastolic and angiography measurements (mean difference 0.2± SD 0.16 mm, P not significant). Conclusions: Our results support the use of navigated 3D-whole-heart CMR to evaluate aortic diameters. Systolic-gated 3D produces larger diameter, especially in the ascending aorta.
KW - Aortic diameter assessment
KW - Aortic wall disease
KW - Cardiovascular magnetic resonance (CMR)
KW - Marfan syndrome
UR - http://www.scopus.com/inward/record.url?scp=85123112658&partnerID=8YFLogxK
U2 - 10.21037/cdt-20-868
DO - 10.21037/cdt-20-868
M3 - Article
AN - SCOPUS:85123112658
SN - 2223-3652
VL - 11
SP - 1389
EP - 1394
JO - Cardiovascular Diagnosis and Therapy
JF - Cardiovascular Diagnosis and Therapy
IS - 6
ER -