TY - JOUR
T1 - A spine frame for intra-operative fixation to increase accuracy in spinal navigation and robotics
AU - Thomale, Ulrich W.
AU - Kneissler, Marc
AU - Hein, Andreas
AU - Maetzig, Martin
AU - Kroppenstedt, Stefan Nikolaus
AU - Lueth, Tim
AU - Woiciechowsky, Christian
PY - 2005/5
Y1 - 2005/5
N2 - Objective: Intra-operative movements due to mechanical ventilation or manipulations are a limiting factor for accurate spinal navigation or robotic-assisted spinal surgery. The purpose of this study was to assess the accuracy of an intra-operative spinal fixation device in an experimental setup. Materials and Methods: We developed a fixation device, attached to the operating table, that combines soft tissue retraction with spinal process fixation. Using a lumbar spine cadaver, tightness of fixation was evaluated using two measurement systems. Accuracy measurements using changes in spatial co-ordinates of implanted reference markers were performed in three segments, following different manipulations of the spine. In addition, for intra-operative movements of the spine during mechanical ventilation, the range of motion was determined in 10 patients during lumbar interbody fusion. Results: The spine frame was easy to use and did not restrict screw insertion. Mean deviations of the markers' in all segments were measured at between 0.35 and 0.8 mm, following pedicle screw insertion and lateral traction. Intra-operative range of motion of the spine was measured with a mean value of 8.7 ± 3.3 mm. Conclusion: Using our spine frame, a rigid fixation following manipulation of the spine was demonstrated. By overcoming the intra-operative movement-dependent inaccuracy, safety in navigated spine surgery and robotic-assisted procedures might be improved.
AB - Objective: Intra-operative movements due to mechanical ventilation or manipulations are a limiting factor for accurate spinal navigation or robotic-assisted spinal surgery. The purpose of this study was to assess the accuracy of an intra-operative spinal fixation device in an experimental setup. Materials and Methods: We developed a fixation device, attached to the operating table, that combines soft tissue retraction with spinal process fixation. Using a lumbar spine cadaver, tightness of fixation was evaluated using two measurement systems. Accuracy measurements using changes in spatial co-ordinates of implanted reference markers were performed in three segments, following different manipulations of the spine. In addition, for intra-operative movements of the spine during mechanical ventilation, the range of motion was determined in 10 patients during lumbar interbody fusion. Results: The spine frame was easy to use and did not restrict screw insertion. Mean deviations of the markers' in all segments were measured at between 0.35 and 0.8 mm, following pedicle screw insertion and lateral traction. Intra-operative range of motion of the spine was measured with a mean value of 8.7 ± 3.3 mm. Conclusion: Using our spine frame, a rigid fixation following manipulation of the spine was demonstrated. By overcoming the intra-operative movement-dependent inaccuracy, safety in navigated spine surgery and robotic-assisted procedures might be improved.
KW - Computer-assisted
KW - Navigation, robotic
KW - Pedicle screw
KW - Spine frame
UR - http://www.scopus.com/inward/record.url?scp=28844506281&partnerID=8YFLogxK
U2 - 10.1080/10929080500229744
DO - 10.1080/10929080500229744
M3 - Article
C2 - 16321912
AN - SCOPUS:28844506281
SN - 1092-9088
VL - 10
SP - 151
EP - 155
JO - Computer Aided Surgery
JF - Computer Aided Surgery
IS - 3
ER -