TY - JOUR
T1 - A Sleeve-Based, Micromotion Avoiding, Retractable and Tear-Opening (SMART) Insertion Tool for Cochlear Implantation
AU - Aebischer, Philipp
AU - Weder, Stefan
AU - Mantokoudis, Georgios
AU - Vischer, Mattheus
AU - Caversaccio, Marco
AU - Wimmer, Wilhelm
N1 - Publisher Copyright:
© 2023 IEEE.
PY - 2023/3/1
Y1 - 2023/3/1
N2 - Objective: In conventional cochlear implantation, the insertion of the electrode array is strongly affected by the local anatomy and human kinematics. Herein, we present a concept for an insertion tool that allows to optimize the insertion trajectory beyond anatomical constraints and stabilizes the electrode array in manual implantation. A novel sleeve-based design allows the instrument to be compliant and potentially protective to intracochlear structures, while a tear-open mechanism allows it to be removed after insertion by simply retracting the tool. Methods: Conventional and tool-guided manual insertions were performed by expert cochlear implant surgeons in an analog temporal bone model that allows to simultaneously record intracochlear pressure, insertion forces and electrode array deformation. Results: Comparison between conventional and tool-guided insertions demonstrate a substantial reduction of maximum insertion forces, force variations, transverse intracochlear electrode array movement, and pressure transients. Conclusion: The presented tool can be utilized in manual cochlear implantation and significantly improves key metrics associated with intracochlear trauma. Significance: The instrument may ultimately help improve hearing outcomes in cochlear implantation. The versatile design may be used in both manual cochlear implantation and motorized and robotic insertion, as well as image-guided surgery.
AB - Objective: In conventional cochlear implantation, the insertion of the electrode array is strongly affected by the local anatomy and human kinematics. Herein, we present a concept for an insertion tool that allows to optimize the insertion trajectory beyond anatomical constraints and stabilizes the electrode array in manual implantation. A novel sleeve-based design allows the instrument to be compliant and potentially protective to intracochlear structures, while a tear-open mechanism allows it to be removed after insertion by simply retracting the tool. Methods: Conventional and tool-guided manual insertions were performed by expert cochlear implant surgeons in an analog temporal bone model that allows to simultaneously record intracochlear pressure, insertion forces and electrode array deformation. Results: Comparison between conventional and tool-guided insertions demonstrate a substantial reduction of maximum insertion forces, force variations, transverse intracochlear electrode array movement, and pressure transients. Conclusion: The presented tool can be utilized in manual cochlear implantation and significantly improves key metrics associated with intracochlear trauma. Significance: The instrument may ultimately help improve hearing outcomes in cochlear implantation. The versatile design may be used in both manual cochlear implantation and motorized and robotic insertion, as well as image-guided surgery.
KW - Cochlear implant
KW - insertion tool
KW - surgical tool
UR - http://www.scopus.com/inward/record.url?scp=85137892859&partnerID=8YFLogxK
U2 - 10.1109/TBME.2022.3204069
DO - 10.1109/TBME.2022.3204069
M3 - Article
C2 - 36063524
AN - SCOPUS:85137892859
SN - 0018-9294
VL - 70
SP - 860
EP - 866
JO - IEEE Transactions on Biomedical Engineering
JF - IEEE Transactions on Biomedical Engineering
IS - 3
ER -