1 μ s broadband frequency sweeping reflectometry for plasma density and fluctuation profile measurements

F. Clairet, C. Bottereau, A. Medvedeva, D. Molina, G. D. Conway, A. Silva, U. Stroth

Research output: Contribution to journalArticlepeer-review

25 Scopus citations

Abstract

Frequency swept reflectometry has reached the symbolic value of 1 μs sweeping time; this performance has been made possible, thanks to an improved control of the ramp voltage driving the frequency source. In parallel, the memory depth of the acquisition system has been upgraded and can provide up to 200 000 signals during a plasma discharge. Additional improvements regarding the trigger delay determination of the acquisition and the voltage ramp linearity required by this ultra-fast technique have been set. While this diagnostic is traditionally dedicated to the plasma electron density profile measurement, such a fast sweeping rate can provide the study of fast plasma events and turbulence with unprecedented time and radial resolution from the edge to the core. Experimental results obtained on ASDEX Upgrade plasmas are presented to demonstrate the performances of the diagnostic.

Original languageEnglish
Article number113506
JournalReview of Scientific Instruments
Volume88
Issue number11
DOIs
StatePublished - 1 Nov 2017