TY - JOUR
T1 - Validation of a Robotic Testbench for Evaluating Biomechanical Effects of Implant Rotation in Total Knee Arthroplasty on a Cadaveric Specimen
AU - Wilhelm, Nikolas
AU - von Deimling, Constantin
AU - Haddadin, Sami
AU - Glowalla, Claudio
AU - Burgkart, Rainer
N1 - Publisher Copyright:
© 2023 by the authors.
PY - 2023/9
Y1 - 2023/9
N2 - In this study, we developed and validated a robotic testbench to investigate the biomechanical compatibility of three total knee arthroplasty (TKA) configurations under different loading conditions, including varus–valgus and internal–external loading across defined flexion angles. The testbench captured force–torque data, position, and quaternion information of the knee joint. A cadaver study was conducted, encompassing a native knee joint assessment and successive TKA testing, featuring femoral component rotations at −5°, 0°, and +5° relative to the transepicondylar axis of the femur. The native knee showed enhanced stability in varus–valgus loading, with the +5° external rotation TKA displaying the smallest deviation, indicating biomechanical compatibility. The robotic testbench consistently demonstrated high precision across all loading conditions. The findings demonstrated that the TKA configuration with a +5° external rotation displayed the minimal mean deviation under internal–external loading, indicating superior joint stability. These results contribute meaningful understanding regarding the influence of different TKA configurations on knee joint biomechanics, potentially influencing surgical planning and implant positioning. We are making the collected dataset available for further biomechanical model development and plan to explore the 6 Degrees of Freedom (DOF) robotic platform for additional biomechanical analysis. This study highlights the versatility and usefulness of the robotic testbench as an instrumental tool for expanding our understanding of knee joint biomechanics.
AB - In this study, we developed and validated a robotic testbench to investigate the biomechanical compatibility of three total knee arthroplasty (TKA) configurations under different loading conditions, including varus–valgus and internal–external loading across defined flexion angles. The testbench captured force–torque data, position, and quaternion information of the knee joint. A cadaver study was conducted, encompassing a native knee joint assessment and successive TKA testing, featuring femoral component rotations at −5°, 0°, and +5° relative to the transepicondylar axis of the femur. The native knee showed enhanced stability in varus–valgus loading, with the +5° external rotation TKA displaying the smallest deviation, indicating biomechanical compatibility. The robotic testbench consistently demonstrated high precision across all loading conditions. The findings demonstrated that the TKA configuration with a +5° external rotation displayed the minimal mean deviation under internal–external loading, indicating superior joint stability. These results contribute meaningful understanding regarding the influence of different TKA configurations on knee joint biomechanics, potentially influencing surgical planning and implant positioning. We are making the collected dataset available for further biomechanical model development and plan to explore the 6 Degrees of Freedom (DOF) robotic platform for additional biomechanical analysis. This study highlights the versatility and usefulness of the robotic testbench as an instrumental tool for expanding our understanding of knee joint biomechanics.
KW - cadaver study
KW - implant alignment
KW - knee joint biomechanics
KW - robotic testbench
KW - total knee arthroplasty
UR - http://www.scopus.com/inward/record.url?scp=85170345613&partnerID=8YFLogxK
U2 - 10.3390/s23177459
DO - 10.3390/s23177459
M3 - Article
C2 - 37687914
AN - SCOPUS:85170345613
SN - 1424-8220
VL - 23
JO - Sensors (Switzerland)
JF - Sensors (Switzerland)
IS - 17
M1 - 7459
ER -