Transport von blutgasproben: Ist die rohrpost sicher?

R. Zanner, N. Moser, M. Blobner, P. B. Luppa

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

9 Zitate (Scopus)

Abstract

Background. Transport of blood gas samples via a pneumatic tube system and subsequent analysis in the central laboratory can reduce costs and errors compared to on-site testing in the operating theatre or the intensive care unit. In this study, a modern pneumatic tube transport system was tested for its usability for this purpose. Methods. A total of 4 consecutive blood gas samples were obtained intraoperatively from 54 different patients and sent to the central laboratory. Of these, 3 samples were transferred using the pneumatic tube system but by different methods and 1 sample was transported personally which served as a reference. The results of sample analysis concerning blood gases, electrolytes and haemoglobin were compared and examined for differences. Results. No statistically significant differences could be determined between the different modes of transportation. Conclusion. Transport of samples for blood gas analysis via a modern pneumatic tube system is safe when samples are correctly prepared.

Titel in ÜbersetzungTransport of blood gas samples: Is the pneumatic tube system safe?
OriginalspracheDeutsch
Seiten (von - bis)1099-1104
Seitenumfang6
FachzeitschriftAnaesthesist
Jahrgang55
Ausgabenummer10
DOIs
PublikationsstatusVeröffentlicht - Okt. 2006

Schlagwörter

  • Artefacts
  • Blood gas analysis
  • Point-of-care testing
  • Samples

Fingerprint

Untersuchen Sie die Forschungsthemen von „Transport von blutgasproben: Ist die rohrpost sicher?“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren