T1 - Successful treatment of a paravalvular leak with balloon cracking and valve-in-valve TAVR
AU - Ruge, Hendrik
AU - Erlebach, Magdalena
AU - Lieberknecht, Eveline
AU - Lange, Rüdiger
N1 - Publisher Copyright:
© 2019 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals, Inc.
PY - 2020/3/1
Y1 - 2020/3/1
N2 - Transcatheter heart valve implantation into degenerated bioprosthetic valves (ViV-THV implantation) has become an established procedure for high risk patients. In general, paravalvular leak (PVL) is a contraindication for valve-in-valve-TAVR (ViV-TAVR). Herein, we report on a 81-year-old patient presenting with acute heart failure for a failing aortic bioprosthesis (Medtronic Mosaic 27 mm). Intraoperative transesophageal echocardiography during urgent ViV-TAVR revealed a PVL previously not detected. After transfemoral implantation of a 26 mm-Evolut-R, balloon-fracturing of the bioprosthetic ring was performed using a 24 mm True Dilatation balloon for treatment of the PVL. Afterward, left ventricular to aortic peak-to-peak pressure gradient measured 2–4mmHg. Transesophageal echocardiography merely revealed trace PVL. Aortic root angiography showed no PVL. At discharge, echocardiography measured a transprosthetic mean gradient of 5mmHg detecting no PVL. Intentional ring-fracturing of an aortic valve prostheses may prove not only to be effective in lowering transvalvular gradients after valve-in-valve-TAVR, but may also be a tool to treat PVL alongside degenerated surgical aortic bioprostheses in certain patients.
AB - Transcatheter heart valve implantation into degenerated bioprosthetic valves (ViV-THV implantation) has become an established procedure for high risk patients. In general, paravalvular leak (PVL) is a contraindication for valve-in-valve-TAVR (ViV-TAVR). Herein, we report on a 81-year-old patient presenting with acute heart failure for a failing aortic bioprosthesis (Medtronic Mosaic 27 mm). Intraoperative transesophageal echocardiography during urgent ViV-TAVR revealed a PVL previously not detected. After transfemoral implantation of a 26 mm-Evolut-R, balloon-fracturing of the bioprosthetic ring was performed using a 24 mm True Dilatation balloon for treatment of the PVL. Afterward, left ventricular to aortic peak-to-peak pressure gradient measured 2–4mmHg. Transesophageal echocardiography merely revealed trace PVL. Aortic root angiography showed no PVL. At discharge, echocardiography measured a transprosthetic mean gradient of 5mmHg detecting no PVL. Intentional ring-fracturing of an aortic valve prostheses may prove not only to be effective in lowering transvalvular gradients after valve-in-valve-TAVR, but may also be a tool to treat PVL alongside degenerated surgical aortic bioprostheses in certain patients.
UR - http://www.scopus.com/inward/record.url?scp=85076100017&partnerID=8YFLogxK
U2 - 10.1002/ccd.28644
DO - 10.1002/ccd.28644
M3 - Article
C2 - 31794136
AN - SCOPUS:85076100017
SN - 1522-1946
VL - 95
SP - 859
EP - 862
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
IS - 4
ER -