Scanning probe microscopy experiments in microgravity

Tanja Drobek, Michael Reiter, Wolfgang M. Heckl

Publikation: Beitrag in FachzeitschriftKonferenzartikelBegutachtung

3 Zitate (Scopus)

Abstract

The scanning probe microscopy setups are small, lightweight and do not require vacuum or high voltage supply. In addition, samples can be investigated directly without further preparation. Therefore, these techniques are well-suited for applications in space, in particular, for operation on the International Space Station (ISS) or for high resolution microscopy on planetary missions. A feasibility study for a scanning tunneling microscopy setup was carried out on a parabolic flight campaign in November 2001 in order to test the technical setup for microgravity applications. With a pocket-size design microscope, a graphite surface was imaged under ambient conditions. Atomic resolution was achieved although the quality of the images was inferior in comparison to laboratory conditions. Improvements for future scanning probe microscopy experiments in microgravity are suggested.

OriginalspracheEnglisch
Seiten (von - bis)3-8
Seitenumfang6
FachzeitschriftApplied Surface Science
Jahrgang238
Ausgabenummer1-4 SPEC. ISS.
DOIs
PublikationsstatusVeröffentlicht - 15 Nov. 2004
Extern publiziertJa
VeranstaltungAPHYS 2003 - Badajoz, Spanien
Dauer: 13 Okt. 200318 Okt. 2003

Fingerprint

Untersuchen Sie die Forschungsthemen von „Scanning probe microscopy experiments in microgravity“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren