Parabolic versus elliptic focusing - Optimization of the focusing design of a cold triple-axis neutron spectrometer by Monte-Carlo simulations

A. C. Komarek, P. Böni, M. Braden

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

15 Zitate (Scopus)

Abstract

We present Monte-Carlo simulations for the focusing design of a novel cold-neutron triple-axis spectrometer to be installed at the end position of the cold guide NL-1 of the research reactor FRM-II in Munich, Germany. Our simulations are of general relevance for the design of triple-axis spectrometers at end positions of neutron guides. Using the McStas program code we performed ray trajectories to compare parabolic and elliptic focusing concepts. In addition the design of the monochromator was optimized concerning crystal size and mosaic spread. The parabolic focusing concept is superior to the elliptic alternative in view of the neutron intensity distribution as a function of energy and divergence. In particular, the elliptical configuration leads to an inhomogeneous divergence distribution.

OriginalspracheEnglisch
Seiten (von - bis)63-72
Seitenumfang10
FachzeitschriftNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Jahrgang647
Ausgabenummer1
DOIs
PublikationsstatusVeröffentlicht - 11 Aug. 2011