TY - JOUR
T1 - Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing?-A biomechanical comparison against miniplates on human specimen
AU - Lang, Jan J.
AU - Bastian, Mirjam
AU - Foehr, Peter
AU - Seebach, Michael
AU - Weitz, Jochen
AU - Von Deimling, Constantin
AU - Schwaiger, Benedikt J.
AU - Micheler, Carina M.
AU - Wilhelm, Nikolas J.
AU - Grosse, Christian U.
AU - Kesting, Marco
AU - Burgkart, Rainer
N1 - Publisher Copyright:
Copyright © 2021 Lang et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2021/6
Y1 - 2021/6
N2 - In this study, topology optimized, patient specific osteosynthesis plates (TOPOS-implants) are evaluated for the mandibular reconstruction using fibula segments. These shape optimized implants are compared to a standard treatment with miniplates (thickness: 1.0 mm, titanium grade 4) in biomechanical testing using human cadaveric specimen. Mandible and fibula of 21 body donors were used. Geometrical models were created based on automated segmentation of CT-scans of all specimens. All reconstructions, including cutting guides for osteotomy as well as TOPOS-implants, were planned using a custom-made software tool. The TOPOS-implants were produced by electron beam melting (thickness: 1.0 mm, titanium grade 5). The fibula-reconstructed mandibles were tested in static and dynamic testing in a multi-axial test system, which can adapt to the donor anatomy and apply side-specific loads. Static testing was used to confirm mechanical similarity between the reconstruction groups. Force-controlled dynamic testing was performed with a sinusoidal loading between 60 and 240 N (reconstructed side: 30% reduction to consider resected muscles) at 5 Hz for up to 5 · 105 cycles. There was a significant difference between the groups for dynamic testing: All TOPOS-implants stayed intact during all cycles, while miniplate failure occurred after 26.4% of the planned loading (1.32 · 105 ± 1.46 · 105 cycles). Bone fracture occurred in both groups (miniplates: n = 3, TOPOS-implants: n = 2). A correlation between bone failure and cortical bone thickness in mandible angle as well as the number of bicortical screws used was demonstrated. For both groups no screw failure was detected. In conclusion, the topology optimized, patient specific implants showed superior fatigue properties compared to miniplates in mandibular reconstruction. Additionally, the patient specific shape comes with intrinsic guiding properties to support the reconstruction process during surgery. This demonstrates that the combination of additive manufacturing and topology optimization can be beneficial for future maxillofacial surgery.
AB - In this study, topology optimized, patient specific osteosynthesis plates (TOPOS-implants) are evaluated for the mandibular reconstruction using fibula segments. These shape optimized implants are compared to a standard treatment with miniplates (thickness: 1.0 mm, titanium grade 4) in biomechanical testing using human cadaveric specimen. Mandible and fibula of 21 body donors were used. Geometrical models were created based on automated segmentation of CT-scans of all specimens. All reconstructions, including cutting guides for osteotomy as well as TOPOS-implants, were planned using a custom-made software tool. The TOPOS-implants were produced by electron beam melting (thickness: 1.0 mm, titanium grade 5). The fibula-reconstructed mandibles were tested in static and dynamic testing in a multi-axial test system, which can adapt to the donor anatomy and apply side-specific loads. Static testing was used to confirm mechanical similarity between the reconstruction groups. Force-controlled dynamic testing was performed with a sinusoidal loading between 60 and 240 N (reconstructed side: 30% reduction to consider resected muscles) at 5 Hz for up to 5 · 105 cycles. There was a significant difference between the groups for dynamic testing: All TOPOS-implants stayed intact during all cycles, while miniplate failure occurred after 26.4% of the planned loading (1.32 · 105 ± 1.46 · 105 cycles). Bone fracture occurred in both groups (miniplates: n = 3, TOPOS-implants: n = 2). A correlation between bone failure and cortical bone thickness in mandible angle as well as the number of bicortical screws used was demonstrated. For both groups no screw failure was detected. In conclusion, the topology optimized, patient specific implants showed superior fatigue properties compared to miniplates in mandibular reconstruction. Additionally, the patient specific shape comes with intrinsic guiding properties to support the reconstruction process during surgery. This demonstrates that the combination of additive manufacturing and topology optimization can be beneficial for future maxillofacial surgery.
UR - http://www.scopus.com/inward/record.url?scp=85107435077&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0253002
DO - 10.1371/journal.pone.0253002
M3 - Article
C2 - 34101755
AN - SCOPUS:85107435077
SN - 1932-6203
VL - 16
JO - PLoS ONE
JF - PLoS ONE
IS - 6 June
M1 - e0253002
ER -