Highly parallel affinity sensor for the detection of environmental contaminants in water

Michael G. Weller, Andreas J. Schuetz, Michael Winklmair, Reinhard Niessner

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

65 Zitate (Scopus)

Abstract

A parallel affinity sensor array (PASA) was implemented based on chemiluminescence labels (peroxidase/luminol) and CCD detection. Reagents like antibodies or haptens have been immobilized on a glass slide, which forms a biochip with an active area of about 1.8cm2. The spot density on the biochip was up to 9 spots per mm2, which corresponds to an array of 1600 spots. Direct and indirect immunoassay formats have been tested successfully. Detection limits down to 20ngl-1 terbutylazine could be reached. The regeneration of the indirect assay chips could be performed more than 100 times. To improve the regenerability of direct assays an enzyme tracer with a cleavable spacer was introduced. The potential of the PASA for analyte identification by pattern recognition and subsequent analyte quantification could be shown. The application of the system is discussed for environmental, pharmaceutical, and other analytical fields. Copyright (C) 1999 Elsevier Science B.V.

OriginalspracheEnglisch
Seiten (von - bis)29-41
Seitenumfang13
FachzeitschriftAnalytica Chimica Acta
Jahrgang393
Ausgabenummer1-3
DOIs
PublikationsstatusVeröffentlicht - 30 Juni 1999

Fingerprint

Untersuchen Sie die Forschungsthemen von „Highly parallel affinity sensor for the detection of environmental contaminants in water“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren