Efficacy and safety of ECG-synchronized pulsatile extracorporeal membrane oxygenation in the clinical setting: The SynCor Trial

Ingo Voigt, Tobias Spangenberg, Tareq Ibrahim, Christian Bradaric, Achim Viertel, Ezequiel Marcelo Tallone, Carsten Skurk, Peter Abel, Jochen Graf, Torsten Rinne, Jürgen Böhm, Alexander Ghanem, Christoph Liebetrau

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

8 Zitate (Scopus)

Abstract

Introduction: Mechanical circulatory support (MCS) devices are increasingly used as a treatment option in resuscitation or in patients with cardiogenic shock (CS). Prophylactic implantation in high-risk percutaneous coronary interventions (HRPCI) is another upcoming indication. The i-cor ECG-synchronized cardiac assist device combines the hemodynamic support of a veno-arterial extracorporeal membrane oxygenation (VA-ECMO) with the ability to generate a pulsatile flow and thus decreasing adverse effects of VA-ECMO on myocardial function. Aim of this study was to obtain data concerning feasibility, safety and outcomes in both indications. Methods: A total of 47 patients (34 HRPCI, 13 CS) were included in nine German centers and participated in this study. Demographic and clinical parameters, procedural as well as follow-up data were prospectively recorded and analyzed. Results: Device implantation and initiation of ECG-synchronized cardiac assist was technical successful in all cases and no failures of the consoles or disposable parts were observed. Furthermore, intended percutaneous coronary interventions and successful weaning from cardiac assist was achieved in 97.1% of HRPCI patients. We observed a 30d-survival of 94.1% in the HRPCI group and 69.2% in the CS group. Main complications in both groups were bleeding events (14.7% HRPCI, 23.1% CS) and critical limb ischemia (2.9% HRPCI, 38.5% CS). Conclusion: The i-cor ECG-synchronized cardiac assist device appears safe and feasible showing clinical outcomes comparable to existing data in the setting of high-risk percutaneous coronary interventions and acute cardiogenic shock. Further prospective trials are warranted to identify optimal patient and interventional characteristics that will benefit most of this novel kind of mechanical circulatory support.

OriginalspracheEnglisch
Seiten (von - bis)387-397
Seitenumfang11
FachzeitschriftArtificial Organs
Jahrgang46
Ausgabenummer3
DOIs
PublikationsstatusVeröffentlicht - März 2022

Fingerprint

Untersuchen Sie die Forschungsthemen von „Efficacy and safety of ECG-synchronized pulsatile extracorporeal membrane oxygenation in the clinical setting: The SynCor Trial“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren