Abstract
Purpose: The purpose of this study was to evaluate the dose characteristic for patient examinations at the first clinical X-ray dark-field chest radiography system and to determine whether the effective patient dose is within a clinically acceptable dose range. Methods: A clinical setup for grating-based dark-field chest radiography was constructed and commissioned, operating at a tube voltage of 70 kVp. Thermoluminescent dosimeter (TLD) measurements were conducted using an anthropomorphic phantom modeling the reference person to obtain a conversion coefficient relating dose area product (DAP) to effective patient dose at the dark-field system. For 92 patients, the DAP values for posterior–anterior measurements were collected at the dark-field system. Using the previously determined conversion coefficient, the effective dose was calculated. Results: A reference person, modeled by an anthropomorphic phantom, receives an effective dose of 35 µSv. For the examined patients, a mean effective dose of 39 µSv was found. Conclusions: The effective dose at the clinical dark-field radiography system, generating both attenuation and dark-field images, is within the range of reported standard dose values for chest radiography.
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 6152-6159 |
Seitenumfang | 8 |
Fachzeitschrift | Medical Physics |
Jahrgang | 48 |
Ausgabenummer | 10 |
DOIs | |
Publikationsstatus | Veröffentlicht - Okt. 2021 |