Abstract
The titanium content of pyrope garnet can be quantified using the intensity of a Raman band at about 830 cm-1 that is normalized to the 363 cm-1 band using a spectrometer-specific calibration using 10 to 15 chromian pyropes from Bohemia, Czech Republic. An accuracy of 0.025 wt% could be achieved for TiO2 contents between 0.17 and 0.67 wt% TiO2 with a Raman spectrometer with a spectral resolution of better than 3.9 cm-1. The technique can be used in petrological and gemmological studies.
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 486-491 |
Seitenumfang | 6 |
Fachzeitschrift | Journal of Raman Spectroscopy |
Jahrgang | 47 |
Ausgabenummer | 4 |
DOIs | |